20
AECG100 | User Manual
Figure 15: Signal Detection
Note:
Because the light intensity of the DUT varies, the "signal strength" may not
be filled. Please adjust the position and find a relatively large signal value.
2.5
Operation Modes
AECG100 can be controlled through PC software operation or
standalone operation.
2.5.1
PC Software Operation
AECG100 must be connected to the computer and the PC
software should be installed properly. AECG100 offers a total
of 4 test modes, including ECG test mode, PPG test mode,
PWTT test mode, and SpO
2
test mode. Auto Sequence test
mode can combine any of the 4. Please see chapter 4 Software
Interface for a detailed operation description.
ECG test mode: AECG100 can conduct single-lead ECG testing
under this test mode. All the ECG-related standards are
supported including a series of single-channel testing, such as
sensitivity, frequency response, input impedance, and so on.
This test mode is compliant with international standards of YY-
0885 and IEC 60601-2-47.