12
AECG100 | User Manual
Figure 2: Set Mode A/B/C
Under standalone operation, the user can press the button
to switch among the modes. There is a total of 3 modes to
choose from, and only the LED light of the chosen mode will
be on. The default mode is Mode A.
Under standalone operation mode, switching modes
require loading the restored parameters while LED flashes
4 times per second. The LED remains on when the restored
parameters are loaded successfully.
Mode A/B/C: The default parameters are for PWTT tests.
Note:
When testing AECG100 main console with different PPG
modules, re-saving Mode A/B/C is required in order to obtain
desired test results.
(14)
Mode Selection Button:
To select amongst Mode A/B/C.
(15)
DB15 Port (Female):
The port on the AECG100 main console to
be connected with various PPG modules.
(16)
DB15 Port (Male):
The port on the PPG module to be connected
with that of (1)AECG100 main console.
(17)
LED & Photodiode (PD):
Receiving the LED signal of DUT and
emitting the LED signal to DUT.
(18)
LED Monitor Port:
This is for the PPG-1R-525 PPG module to
connect to the oscilloscope and measure the output voltage of