background image

Operation Manual 

BTpro Botanical 

Diaphragm

 Pump

s

BTpro170 - BTpro100 UL

 

201

9

-

11

-

0

1

 

8-Headed BTpro170 

Models: 

7040, 7042

4

-Headed BTpro1

0

Models: 

70

3

0, 70

3

2

Summary of Contents for 7030

Page 1: ...Operation Manual BTpro Botanical Diaphragm Pumps BTpro170 BTpro100 UL 2019 11 01 8 Headed BTpro170 Models 7040 7042 4 Headed BTpro100 Models 7030 7032 ...

Page 2: ... reserved by Co Gardner Denver Thomas GmbH We reserve the right to make changes and amendments Gardner Denver Thomas GmbH Am Vogelherd 20 98693 Ilmenau Germany For USA Canada and other Americas Gardner Denver Thomas Inc 1601 Feehanville Drive Suite 600 Mount Prospect IL 60056 USA T 1 847 676 8800 F 1 847 677 8606 welch na gardnerdenver com www welchvacuum com T 49 3677 604 0 F 49 3677 604 131 welc...

Page 3: ...epair 3 Keep the cord away from heated surfaces 4 Never use while sleeping or drowsy 5 Do not operate this product where oxygen is being administered 6 This unit is thermally protected See page 2 Thermal Switch Protection for full details 7 Always disconnect power source before servicing 8 Wear safety glasses or goggles when operating this product 9 Use only in well ventilated areas 10 All electri...

Page 4: ...rotection will continue to stay open Operation The intended use for this vacuum pump is for general physical and chemical laboratory applications requiring vacuum ranging from atmospheric pressure down to an ultimate vacuum of 8 mbar 1 The pump is equipped with an ON OFF I O power switch 2 Gas Ballast A gas ballast valve A is provided on the pump When condensable vapors are pumped they may be comp...

Page 5: ...to tighten the pressure disc 8 the diaphragm 7 and the strain washer 6 with the correct torque of 5 6 Nm 3 Bring the piston rod see Fig 4 6 and the diaphragm 7 into the central position 4 Replace the pump head 5 5 Insert the valves 3 and the O rings 4 Ensure that they are lying completely flat Do not insert the burred side facing the sealing surface Align the connection head flush with the pin 6 T...

Page 6: ...ce kit consists of Designation Kit Order no 402043 US Kit Order no 402042 01 US for Model BTpro170 for Model BTpro100 Piece Order no Piece Order no O Ring ø 12 x 2 36 829217 3 8 829217 3 O Ring ø 25 x 2 16 829250 1 8 829250 1 Valve 16 400656 6 400656 Valve with hole 2 400656 4 Diaphragm 8 400732 4 400732 4 ...

Page 7: ...this information Parameter Unit Diaphragm pump BTpro170 two stage Free Air Displacement 60 Hz CFM 6 0 L min 170 Ultimate pressure mbar 8 Max inlet pressure bar 1 Max outlet pressure 1 IN Intake 1 and EX Exhaust 2 connections standard Small flange DN 16 KF option Hose nozzle DN 8 for hose inside diameter 8 mm enclosed Ambient temperature C 10 to 40 Max operating gas temperature 40 Bearing maintenan...

Page 8: ...Spare Parts Overview Spare parts view Basic structure BTpro170 Fig 4 Exploded view Basic structure Diaphragm pumps 8 headed 6 ...

Page 9: ...24949 3 8 Mass balance 2 400678 1 400678 1 9 Alternating current motor 1 826390 5 826390 7 10 Spacer distance Foot 4 410437 410437 11 Foot 1 410435 410435 12 Air duct 2 410438 410438 13 Rubber metal pad 4 829141 2 829141 2 14 Handle 2 828670 828670 15 Spacer for fan 4 103521 103521 16 AC axial fan 1 829819 3 829819 5 17 Time controlling VAPU 0003 1 825681 825681 1 18 Motor operating condenser 20µF...

Page 10: ... 35 Intake exhaust port PP DN16KF 2 710116 710116 36 Straight threaded joint with seal edge PVDF 10 30 829931 829931 37 Threaded elbow joint PVDF 10 M 12 x 1 38 Threaded elbow joint PVDF 10 39 Threaded elbow joint with attitude taps PVDF 10 2 829913 1 829913 1 40 T screw connection PP 10 10 10 41 O Ring EPDM ø 12 x 2 34 829217 3 829217 3 42 Check valve PVDF 8 1 829909 829909 43 Gas ballast valve 1...

Page 11: ...sume any obligation or liability in connection with this information Parameter Unit Diaphragm pump BTpro100 two stage Free Air Displacement 60 Hz CFM 3 5 L min 100 Ultimate pressure mbar 8 Max inlet pressure bar 1 Max outlet pressure 1 IN Intake 1 and EX Exhaust 2 connections standard Small flange DN 16 KF option Hose nozzle DN 8 for hose inside diameter 8 mm enclosed Ambient temperature C 10 to 4...

Page 12: ...Spare Parts Overview Fig 6 Exploded view BTpro100 Spare parts view Vacuum connection BTpro100 10 ...

Page 13: ...PTFE insert 4 400902 400902 17 Pressure washer 4 400680 400680 18 Tightening washer 4 400707 400707 19 Diaphragm 4 400732 400732 20 Valve 8 400656 400656 Valve with hole 2 400656 4 400656 4 21 Manifold 5 2 400917 01 400917 01 22 Manifold 2 2 400903 400903 Manifold 4 1 400911 400911 23 Hose nozzle PP DN 8 2x enclosed 2 710798 710798 24 Straight threaded joint with seal edge PVDF 10 6 829931 829931 ...

Page 14: ...n completed and returned 2 The product is returned to the factory or other designated service centers freight prepaid 3 The product in our judgment is defective through no action or fault of the user If the product has become defective through misuse abuse or alteration repairs will be billed regardless of the age of the product In this event an estimate of the repair costs will be submitted and a...