3. Squeeze the locking tab to disconnect the battery and remove the battery.
Reassembly notes
WARNING
Defective batteries can damage the device. If the battery shows any
signs of damage or cracking, it must be replaced immediately and only with a
battery provided by Welch Allyn.
Kit items:
•
BATT22, BATTERY PACK
•
107212, SRV KIT, BATTERY FOAM
NOTE
Reassembly is the reverse of disassembly.
Remove the back housing
1. Remove the 9 Phillips screws on the back of the device.
42 Disassembly and repair