Reset the device
Reset the device only when the device becomes unresponsive. Patient data and configuration
settings will be cleared from device memory.
1. Press and hold , located on the lower-left corner of the device.
2. If there is a prompt with options to Power down, Sleep, or Cancel, continue to press for
several seconds.
The device powers down. Patient data and configuration settings are cleared from device
memory.
3. Press to power up the device.
Service manual
Power-up sequence 29