USING RSCRIBE
Page | 44
Recording an ECG
RScribe can acquire single or multiple ECGs in real time, retrospectively, or in a timed sequence.
Prepare the patient for best conductivity between the skin surface and the electrode. Place the electrodes on the
patient according to lead placement guidelines. Ensure that the proper lead wire is firmly and correctly attached to
each electrode.
Select an existing order from the MWL tab, search for the patient from the Patients tab, or manually enter patient
demographics.
The real-time ECG display can be exited with selection of the Done
button to return to the
Start a Resting
Exam
display at any time. The user is prompted to select
Yes
or
No
when an ECG has not been acquired.
To acquire a real-time ECG:
Select the ECG button on the display to capture Best 10
or Last 10
12-lead ECG data.
o
BEST 10 captures the best quality, 10-second ECG accumulated within the full disclosure
o
LAST 10 captures the most recently acquired 10-second ECG
Or, select the
ECG
button
on the WAM™ or the AM12™ that will capture either the BEST or
LAST 10-seconds of acquired ECG dependent on RScribe configurations.
Preview Acquisition Screen
The ECG preview screen appears with an ECG report exactly as it will save and print upon exiting this display.
Select the
Save
button to save the ECG as acquired, exit, and return to the
Start a Resting
Exam
display.
Select the
Delete
button to discard the ECG and return to the real-time ECG display to capture
another ECG.
Select the
Review
button to save the ECG as reviewed, exit, and return to the
Start a Resting
Exam
display. The user is prompted to enter the reviewer’s name prior to exit.
Select the
Sign
button to save the ECG as signed, exit, and return to the
Start a Resting
Exam
display. The user is prompted to enter the signer’s name prior to exit. When
Legal Signature
has
been defined in the workflow configuration settings, the signer is prompted to enter the user name and
password.
Select the
Print ECG