GENERAL CARE
Precautions
Turn off the device before inspecting or cleaning.
Do not immerse the device in water.
Do not use organic solvents, ammonia based solutions, alcohol, or abrasive cleaning agents which may
damage equipment surfaces.
Inspection
Inspect your equipment daily prior to operation. If you notice anything that requires repair, contact an authorized
service person to make the repairs.
Verify that all cords and connectors are securely seated.
Check the case and chassis for any visible damage.
Inspect cords and connectors for any visible damage.
Inspect keys and controls for proper function and appearance.
Cleaning and Disinfecting
1.
Disconnect the power source. Remove cables and lead wires from device before cleaning.
2.
For general cleaning of cables and lead wires, use a soft, lint-free cloth lightly moistened with a mild soap
and water solution. Wipe and air dry.
3.
For disinfecting the exterior surfaces of the device, patient acquisition module, cables, and lead wires, wipe
exterior using:
Clorox Healthcare® Bleach Germicidal Wipes (use according to instructions on product label),
or
a soft, lint-free cloth with a solution of Sodium Hypochlorite (10% household bleach and water solution)
minimum 1:500 dilution (minimum 100 ppm free chlorine) and maximum 1:10 dilution as recommended
by the APIC Guidelines for Selection and Use of Disinfectants
WARNING: Prevent liquid from penetrating the device and do not attempt to clean/disinfect
the device or patient cables by submerging into a liquid, autoclaving, or steam cleaning. Never expose
cables to strong ultra-violet radiation. Do not sterilize the device or ECG lead wires with Ethylene
Oxide (EtO) gas.
WARNING: Use of unspecified cleaning/disinfecting agents or failure to follow recommended procedures
could result in increased risk of harm to users, patients and bystanders, or damage to the device.
Disposal
This product and its accessories must be disposed of according to local laws and regulations. Do not dispose of this
product as unsorted municipal waste. For more specific disposal information see
www.welchallyn.com/weee
.