Directions for Use
Specifications
25
Electromagnetic Compatibility
WARNING
AC cables other than those specified by manufacturer may result in
increased EMISSIONS or decreased IMMUNITY.
Electromagnetic Emissions
The Welch Allyn ProXenon 350 Surgical Illuminator is intended for use in the electromagnetic environment specified
below. The customer or user of the Welch Allyn ProXenon 350 Surgical Illuminator should assure that it is used in such
an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The Welch Allyn ProXenon 350 Surgical Illuminator uses RF energy only
for its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The Welch Allyn ProXenon 350 Surgical Illuminator is suitable for use in
all establishments, other than domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies