58
Functional verification
Welch Allyn Propaq LT Vital Signs Monitor
Charge time (optional procedure)
A normally discharged battery pack typically requires 3 hours (4 hours maximum) to fully
charge, whether the monitor is on or off during charging.
1.
Plug the monitor with the discharged battery into a powered cradle, in an environment
with temperature in the range 0
C to 40
C.
2.
Note the time required for the status indicator on the cradle to change from green to
off. This may be timed during
System tests
With the battery fully charged:
1.
Power-on the monitor, observing the power-up sequence to verify that all three LED
indicators light and that all three audio tones sound.
2.
Start a new patient.
3.
Highlight
and press
.
4.
Navigate to
Setup > Service > Voltages/Show Info
and press
.
5.
Verify that the displayed (measured) values for each supply are within the ranges
shown on the screen.
6.
Navigate to
Service > Display > Pixel Test
and press
twice.
7.
Using the navigation arrows, cycle through the six colors on each screen half,
verifying that all pixels are displayed in all colors and with uniform accuracy and
brightness.
8.
Press
to exit the Pixel test.
9.
Go to the
Buttons
test and press
twice.
10. Push each button and verify that all button presses are shown.
11. Press
and
simultaneously to exit the Buttons test.
12. Navigate to
Service > Battery > Battery Info
.
13. Verify that the charger state is
8
.
14. Exit the Service menu.
ECG
1.
Connect a 5-lead ECG cable between the monitor and the ECG simulator.
2.
Set the simulator for
Lead II
,
normal sinus
,
80 bpm
,
1 mv
.
3.
Press
as needed to view a waveform display.
4.
Highlight ‘
II
’ and press
to the display the Waveform Source drop-down menu.
Note
A normally discharged battery pack is one on which a monitor has recently run for
more than 5 minutes just before a low-battery shutdown.
Summary of Contents for Propaq LT 802LT0N
Page 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Page 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Page 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Page 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Page 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Page 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...