Service Manual
Repair Procedures
85
Opening the SpO
2
Module
1.
Separate the SpO
2
module from the monitor (
“Removing the Dual Battery in SpO2
2.
Two tabs secure the SpO
2
module rear panel to the SpO
2
module chassis. Using a
small screwdriver or your finger, press on each of the tabs inside the SpO
2
module
chassis while applying pressure to separate the rear panel from the chassis, as shown
here. (When separating the rear panel from the chassis, make sure the side panels
remain seated in the rear panel.)
3.
Once the tabs allow the rear panel to be removed, separate the rear panel from the
chassis.
The rear panel is sealed to the chassis with rubber O-ring gaskets. These gaskets can
be reused when re-assembling the SpO
2
module.
When closing the SpO
2
module, verify no O-ring gasket slips out of its channel and
becomes pinched between the rear panel and the chassis.
SpO2 Module
Rear Panel
Release
Tabs
Continuous
O-Ring Gasket
0696-53
Rear
Panel
O-Ring
Cord
Release
Tab (2)
SpO2
Board
SpO2
Connector
Panel
Chassis
Speaker
SCP Board
Summary of Contents for Propaq Encore
Page 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Page 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Page 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Page 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Page 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Page 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...