Service Manual
Repair Procedures
81
Replacing the Expansion Module Front Panel
The front panel contains the printer buttons. Like the rear panel, the front is sealed to the
chassis with a rubber O-ring gasket. This gasket can be re-used when replacing the front
panel.
The expansion module front panel is attached to the chassis with four screws. Use the
following procedure to remove the front panel:
1.
Remove the four screws (as shown below) holding the front panel.
2.
For expansion module options with sidestream CO
2
(226, 227, 228, or 229),
disconnect the cables from J1 and J3 on the SSP board.
3.
Separate the front panel from the chassis, disconnecting the front panel cable from
the expansion module button board.
A rubber O-ring gasket seals the front panel to the chassis. This gasket should be
saved for re-installing on the front panel.
When installing the expansion module front panel:
•
Verify the O-ring gasket is in place and not pinched.
•
Torque screws to 6.65in-lbs to 7.35in-lbs (.75N-m to .83N-m)
Note
To replace the sidestream CO2 assemblies, see
0696-48
Summary of Contents for Propaq Encore
Page 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Page 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Page 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Page 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Page 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Page 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...