Welch Allyn ProBP 2400 digital blood pressure device

 

Guidance and manufacturer’s declaration – electromagnetic immunity –  

for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING 

Guidance and manufacturer’s declaration – electromagnetic immunity

The ProBP 2400 Digital Blood Pressure Device is intended for use in the electromagnetic environment specified below. The 

customer or the user of the ProBP 2400 Digital Blood Pressure Device should assure that it is used in such an environment.

Immunity test

IEC 60601 test 

level

Compliance 

level

Electromagnetic environment - guidance

Conducted RF 

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms 

 

150 kHz to 80 

MHz

3 V/m

80 MHz to 2.5 GHz

3V 

3 V/m

Portable and mobile RF communications equipment 

should be used no closer to any part of the ProBP 

2400 Digital Blood Pressure Device, including cables, 

than the recommended separation distance calculated 

from the equation applicable to the frequency of the 

transmitter.

Recommended separation distance

P

V

d

]

5

.

3

[

1

=

P

E

d

]

5

.

3

[

1

=

   

80 MHz to 800 MHz

P

E

d

]

7

[

1

=

   

800 MHz to 2.5 GHz

where 

p

 is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 

manufacturer and 

d

 is the recommended separation 

distance in metres (m).

b

Field strengths from fixed RF transmitters, as 

determined by an electromagnetic site survey,

should 

be less than the compliance level in each frequency 

range.

b

Interference may occur in the vicinity of equipment 

marked with the following symbol:

NOTE 1  At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2  These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures, 

objects and people.

a

      Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 

amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 

electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field 

strength in the location in which the ProBP 2400 Digital Blood Pressure Device   is used exceeds the applicable RF compliance level 

above, the ProBP 2400 Digital Blood Pressure Device   should be observed to verify normal operation. If abnormal performance is 

observed, additional measures may be necessary, such as re orienting or relocating the ProBP 2400 Digital Blood Pressure Device  .

b

      Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m. 

Recommended separation distances between portable and mobile 

RF communications equipment and the EQUIPMENT or SYSTEM - 

for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING 

 

Recommended separation distances between

portable and mobile RF communications equipment and the ProBP 2400 Digital Blood Pressure 

Device   

The ProBP 2400 Digital Blood Pressure Device is intended for use in an electromagnetic environment 
in which radiated RF disturbances are controlled. The customer or the user of the ProBP 2400 Digital 
Blood Pressure Device    can help prevent electromagnetic interference by maintaining a minimum 
distance between portable and mobile RF communications equipment (transmitters) and the ProBP 
2400 Digital Blood Pressure Device    as recommended below, according to the maximum output 
power of the communications equipment

Separation distance according to frequency of transmitter

m

Rated maximum 

output of transmitter 

 

W

150 kHz to 80 MHz 

P

V

d

]

5

.

3

[

1

=

80 MHz to 800 MHz

P

E

d

]

5

.

3

[

1

=

800 MHz to 2.5 GHz

P

E

d

]

7

[

1

=

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above the recommended 
separation distance d in metres (m) can be estimated using the equation applicable to 
the frequency of the transmitter, where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency 
range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation 
is affected by absorption and reflection from structures, objects and people.

Summary of Contents for ProBP 2400

Page 1: ...digitalbloodpressuredevice EN 1 ProBP 2400 Digital Blood Pressure Device Directions for use AdvancingFrontlineCare poweredby CAUTIONFederallawrestrictsthisdevicetosalebyorontheorderofaphysicianorlicensedhealthcareprovider ...

Page 2: ...detectstheappearanceofirregularheartbeatduringmeasurement andgivesawarningsignalwiththereading oncetheirregularheartbeatisdetected Thedevicecanaccuratelymeasurebloodpressureinpregnantpatientsincludingthosewithknownorsuspected preeclampsia TheProBP2400combinestheadvantagesofanautomaticbloodpressuredeviceandauscultatorysphygmomanometer It isdesignedtoprovideconvenient accurateandreliableofficebloodp...

Page 3: ...ent JHypertens2005 23 697 701 E AsmarR BeilinL ImaiY etal EuropeanSocietyofHypertensionrecommendationsforconventional ambulatory andhomebloodpressuremeasurement EuropeanSocietyofHypertensionWorkingGrouponBloodPressureMonitoring JHypertens2005 23 697 701 2003 21 1983 1992 Pregnancy Approximately20 ofwomendevelophypertensionduringpregnancy Pre eclampsiaorToxemia whichmayaffect pregnancy Pre eclampsi...

Page 4: ...4 ...

Page 5: ...ittingthecuffproperly 16 17 TakingmeasurementsinSingleReading 1x ManualandThree ReadingAverage Mode 1x Mode 18 Manual Mode 19 20 Mode 21 SpecialFunctions MAP MeanArterialPressure 22 Irregularheartbeatdetectorin 1x Mode 22 Settingmaximuminflationpressure 23 Takingfewerthanthreemeasurements 24 Skippingthecountdowntime 24 Settingmeasurementintervaltimesin Mode 25 Viewingthestoredvalues 26 Appendix Re...

Page 6: ...ultoperatinginstructions directionsforuse DFU Acopyofthe DFUisavailableonthiswebsite AprintedcopyoftheDFUcanbe orderedfromWelchAllynfordeliverywithin7calendardays F Helpfulnotes Shipping storing andenvironmentsymbols Fragile handlewithcare Recovery Recyclable Recycletheproductseparatefromotherdisposables Keepdry Controlsymbols Poweron Poweroff Start Stop Memory Adjustmeasurementintervaltimes Adjus...

Page 7: ...surecuffandother accessoriesforfrayingorotherdamage Replaceasnecessary WARNINGInaccuratemeasurementrisk Donotusethedeviceonpatientswho areconnectedtoheart lungmachines WARNINGElectricshockhazard Donotopenthedeviceorattemptrepairs There arenouser serviceablepartsinsideProBP2400otherthanbatteryreplacement Onlyperformroutinecleaningandmaintenanceproceduresspecifically describedinthisdirectionsforuse ...

Page 8: ...adtoheatgeneration smoke bursting orfire WARNINGDonotdisassemble modify orsolderthebattery WARNINGForproperpatientelectricalisolationandbatterycharging useonlythe providedexternalpowersupplytochargethedevice WARNINGElectricshockhazard Beforecleaningthedevice disconnectthepower cordfromthepowersourceandthedevice WARNINGTakecaretopreventwaterorotherfluidfromenteringanyconnectors onthedevice Shouldth...

Page 9: ...onWelchAllyn bloodpressuretubing Usingtheseconnectorsonbloodpressurecufftubing createstheriskofmistakenlyconnectingthistubingtoapatient sintravenousline andintroducingairintothepatient scirculatorysystem WARNINGNIBPmeasurementsmaybeinaccurateinthepresenceofexcessive motionartefact Minimizeextremityandcuffmotionduringbloodpressurereadings WARNINGThepositionandphysiologicconditionofthesubjectcanaffe...

Page 10: ...10 3 sec 1x 3x 3 sec On OffSwitch PowerAdapterSocket BatteryCompartment CuffSocket Recharge Indicator Display M Button Memory Mode Switch Start Stop Button Productdescription Nameofparts ...

Page 11: ...e Pulse Rate Pulse Detected Mode Selection Single Reading Manual Three Reading Average Power Low Battery Unit of Measure Auto Inflation Maximum Inflation Pressure Rest Count Down Irregular Heart Beat IHB Power LowBattery ModeSelection SingleReading Manual ThreeReadingAverage UnitofMeasure MaximumInflationPressure Rest CountDown AutoInflation Memory SystolicValue DiastolicValue MeanArterialPressure...

Page 12: ... Version 1 01 DESCRIPTION PART NO TITLE AW CD LABEL DFU PROBP 3400 ML 1 03 DRAWING NO SHEET 1 OF 1 DATE 2014 12 24 DRW BY Microlife SCALE 1 1 VER CHANGE NUMBER APVD DESCRIPTION OF CHANGE A RELEASE TO PRODUCTION D D SEE SAP DIR FOR CHANGE NUMBER APPROVER NAME AND DATE OF APPROVAL 408263 CD DFU PROBP 2400 ML 1 01 103701 DFU PROBP 2400 NOTES 1 USE CMYK COLOR MODE AND WHITE FLOOD FOR PRINTING ONTO THE...

Page 13: ...epoweradapter Selectaproperpowerplugandattachtothepoweradapteras shownbelow Chargethebatterycompletely Whenusingforthefirsttime chargethebatteryuntilthe rechargeindicatoronthedeviceturnstogreen F Refertopage27forthesectionof Usingapoweradapter F Refertopage27forthesection RechargeableBattery ...

Page 14: ...ReleasetheMButtonwhenbacklightilluminates 4 PresstheStart StopButtontoselectthepreferred pressureunit mmHgorkPa 5 PressMButtontoconfirmtheselection Units MAPor PAM 1 mmHg MAP 2 mmHg PAM 3 kPa MAP F MAPistheabbreviationofMeanArterialPressure F PAMorMAPischosendependingonlanguagepreference Initialsetup cont F Factorydefaultsetting ...

Page 15: ...andlargeadultlongcuffsarenotavailableintheUSA Checkyourlocalrepresentativeforavailability Reusable ProBP 2400 Blood Pressure Cuff 14 22cm CHILD www welchallyn com Reusable ProBP 2400 Blood Pressure Cuff 32 52cm LARGE ADULT LONG www welchallyn com LATEX LATEX PVC DEHP LATEX PVC DEHP Reusable ProBP 2400 Blood Pressure Cuff 32 42cm LARGE ADULT www welchallyn com LATEX LATEX Reusable ProBP 2400 Blood ...

Page 16: ...d Pressure Cuff 22 32cm ADULT 3 Wrapandtightenthecuffaroundthearm 4 Leavefreespacewiththesizeof2fingersbetween thearmofthepatientandthecuff Excessivetightness maycausevenouscongestionanddiscolorationof thelimb Ifthecuffiswrappedtooloosely itcannot beinflatedproperlyandthemeasuredvaluesmaybe inaccurate Removeallclothingcoveringorconstricting themeasurementarm Clothingmayinterferewith measurementacc...

Page 17: ...00 Blood Pressure Cuff 22 32cm ADULT www welchallyn com LATEX 5 Cuffsthatdonotfitproperlymayleadtoinaccurate readings Useadifferentsizecuffiftherangeindexatthe endofthecuffdoesnotfallintotherangespecifiedbythe rangestripes F Therangeindexofthecuff shouldfallintothisrange ...

Page 18: ...detheswitchtoselect1x standardsinglemeasurement Manualor Mode automaticthreereadingaverage Turnonthepower TurnonthedevicebypressingtheOn OffSwitchattheback ofthedevicetotheONposition Connectthecufftothedevice Connectthecufftothedevicebyinsertingthecuffconnector intothecuffconnectorsocket 3 sec 1x 3x 3 sec 1x 3x 1x 3x 1x 3x FRefertopage22forthesectionof Setmaximuminflation pressure ...

Page 19: ...digitalbloodpressuredevice 1x Mode standardsinglemeasurement Select 1x Modethenpressthe Start Stop Button toperformasinglebloodpressuremeasurement The measurementreadingisdisplayedandsavedafterthe measurement 1x 3x 1x 3x 1x 3x ...

Page 20: ... P 2 4 0 0 B l o o d P r e s s u r e C u f f 2 2 3 2 c m A D U L T Takingmeasurementsin1x Manualand Mode cont R eu sa b le P ro B P 2 4 0 0 B lo o d P re ss u re C u ff 22 3 2c m A D U LT Startinflation PresstheStart StopButtontostartinflation ofthecuff Whenmaximuminflationpressureisreached theProBP 2400willautomaticallybeginlineardeflationatarateof 3mmHg sec Re inflate PushandHoldtheMButtonduring...

Page 21: ...beforethefirstmeasurementissetat60 seconds Theintervaltimesbetweenmeasurementsisuseradjustableto 15 30 45or60seconds Theaveragemeasurementreadingis displayedandsavedafterthemeasurementsarecomplete 15 30 45 or 60 sec 15 30 45 or 60 sec 1 2 3 60 sec 1x 3x 1x 3x 1x 3x Whenthecuffpressurereaches20mmHgduringthe deflationcycle theremainingpressureisventedandthe ProBP2400goesintotheStandbymode Pushthe St...

Page 22: ...alue Irregularheartbeatdetectorin 1x Mode Thedevicedetectsirregularheartbeatin 1x Mode Theirregularheartbeatsymbolshowsupifduringa bloodpressuremeasurementtheheartratehasvaried bymorethan25 Incaseofanirregularheartbeatthe bloodpressuremeasurementmightbeaffected Itis recommendedtorepeatthemeasurementortochoose the Manual Modeforverification FTheMeanArterialPressure MAPisdeterminedfrom themaximumpea...

Page 23: ...UTO atthebottomofthelist 3 PresstheMButtontoconfirmtheselectedvalue Pushandholdfor3 seconds 3 sec 3 sec Ifthemaximuminflationpressureselected orthe maximuminflationpressureasdeterminedwhilein AUTOmode isnotadequatetodeterminesystolic pressure thedevicewillreinflatetoapressure30 mmHghigherthanthepreviousinflationpressureand attemptanothercycle Thiscanberepeatedincreasing themaximuminflationby30mmHg...

Page 24: ...de Thecountdownbeforeandbetweenmeasurementsin ModecanbeskippedbypressingtheStart StopButton When theStart StopButtonispressedduringcountdown thedevice willimmediatelybeginthenextmeasurement FThedevicecanbesetin Standby ModebypressingtheStart StopButtonaftercompletionofmeasurements Thedevicewill automaticallyswitchto Standby Modeifleftunattendedfor1 minute Cancelremaining measurementsat anytimeduri...

Page 25: ...efaultmeasurementintervaltimeis60seconds The intervaltimescanbesetas15 30 45or60seconds 1 PressandholdtheStart StopButtonfor3seconds Pressandholdfor3 seconds 2 PresstheMButtontoadjustthemeasurementinterval time thenpresstheStart StopButtontoconfirm The devicewillgobackto Standby 15 30 60 45 3 sec ...

Page 26: ...ngtheMButtonto reviewindividualmeasurements Viewingstoredvalues Thedeviceonlystoresbloodpressurevaluesofthelast measurementprocedurein 1x and Mode Pressthe MButtontoreviewthestoredreadingswhenthedeviceis in Standby mode FThedevicestoresonlythelastmeasurementcompletedin 1x Modeandthelastthreemeasurementscompletedin Mode ...

Page 27: ...1 Plugtheadaptercableintothepowersocketofthedevice 2 Plugtheadapterplugintothewallsocket Thebattery willberechargedaslongasthedeviceisattached toan ACpowersource Afterthebatteryisfully recharged the chargingwillstop Nobatterypowerwillbeusedaslong astheadapterispluggedin Thebatteryhastoremainin theProBP2400alsowhenusingtheACpower 3 Ifthebatterystartslosing capacity contactyourlocal dealerforreplace...

Page 28: ... Rechargetherechargeablebatterybyplugginginthe powersupply Cuffdoesnotinflateproperly Looseconnectionofthetube Makesurethetubeofthecuffissecurelyconnectedtothe device Leakageofthetube bladder Checkforcracksonthetubeorthebladder Replacethe bloodpressurecuffifnecessary Noresultdisplayedafter measurements DeviceisinManualMode Switchto 1x or Modeandrepeatthe measurements ...

Page 29: ...r isdisplayed Error Description Potentialcauseand remedy Err1 Signaltoo weak Thepulsesignalsonthe cuffaretooweak Re positionthecuffandrepeat themeasurement Err2 Errorsignal Duringthemeasurement errorsignalsweredetected bythecuffcaused for instance bymovementor muscletension Repeat themeasurementkeeping patient sarmstill Errormessages ...

Page 30: ...pulseistoolow less than40beatsperminute Repeatthemeasurement Err3 Nopressurein thecuff Anadequatepressure cannotbegeneratedin thecuff Aleakmayhave occurred Replacethe bloodpressurecuffif necessary Repeatthe measurement Err5 Novalid results Themeasuringsignalsare inaccuratethereforeno resultcan bedisplayed SwitchtoManual Modeanddetermine bloodpressurewith astethoscopeusing traditionalKorotkoff meth...

Page 31: ...otection Thisdevicemaybeusedonlyforthepurposedescribed inthisdirectionsforusemanual Thedevicecomprisesof sensitivecomponentsandmustbetreatedwithcare The manufacturercannotbeheldliablefordamagecausedby incorrectapplication Observethestorageandoperatingconditionsdescribedin the Technicalspecifications sectionofthisdirectionsfor usemanual Onlyactivatethepumpwhenthecuffisconnectedto thedevice Donotuse...

Page 32: ...ffcoverin75 C 167 F hot waterfor30minutes Airdrythecuff DONOTironthecuffcover Reusable ProBP 2400 Blood Pressure Cuff 22 32cm ADULT www welchallyn com LATEX Devicecare Useasoftclothwithoneofthefollowingrecommended cleaningsolutionstowipetheexteriorofthedevice 1 Mildsoapandwater 2 Hydrogenperoxidesolution 3 dilutedwithwater 3 Sodiumhypochloritesolution 1 10dilutionof householdchloridebleachinwater ...

Page 33: ...curacyevery2 yearsoraftermechanicalimpact e g beendropped 1 Setupforaccuracytest 2 Pushandhold Start Stop Buttonandturnthepoweron thenreleasethe Start Stop Button Waituntil CA 0 then 0 0 isdisplayed 350 500 ml Vessel 350 500 ml Vessel Air release valve Calibrated reference manometer ProBP 2400 T pieces Squeeze bulb or Pressure source ...

Page 34: ...ssuretonearly200mmHg Comparethe pressuredisplayedonthescreensofthedeviceand thereferencemanometer Forexample the 201 0 displayedonthedevicestandsfor 201 0mmHg 5 Pumpthepressuretonearly300mmHg Comparethe pressuredisplayedonthescreensofthedeviceand thereferencemanometer Forexample the 301 2 displayedonthedevicestandsfor 301 2mmHg 6 Ifthedifferencebetweenthedeviceandthereference manometeratanycalibra...

Page 35: ...servestherighttoaltertechnicalspecificationswithoutpriorwrittennotice ProBP 2400 ProBP 2400 Digital Blood Pressure Monitor www welchallyn com BOM number Color CMYK 901096 2400 DIGITAL BLOOD PRESSURE DEVICE ML Distributed by Welch Allyn Inc 4341 State Street Road Skaeateles Falls NY 13153 Microlife Corporation 9F 431 RuiGuang Road NeiHu Taipei 11492 Taiwan R O C Microlife AG Espenstrasse 139 9443 W...

Page 36: ...vice should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrostatic transient burst ...

Page 37: ...ectromagnetic site survey should be considered If the measured field strength in the location in which the ProBP 2400 Digital Blood Pressure Device is used exceeds the applicable RF compliance level above the ProBP 2400 Digital Blood Pressure Device should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocat...

Page 38: ...sedby 1 handlingduringshipping 2 useormaintenancecontraryto labeledinstructions 3 alterationorrepairbyanyonenotauthorizedbyWelchAllyn and4 accidents Theproductwarrantyisalsosubjecttothefollowingtermsandlimitations ShippingcosttoreturnadevicetoaWelchAllynServicecenterisnotincluded AservicenotificationnumbermustbeobtainedfromWelchAllynpriortoreturninganyproductsoraccessoriesto WelchAllyn sdesignated...

Page 39: ...bed inthispublication thepurchaser oftheproductispermittedtocopythispublication forinternaldistributiononly fromthemediaprovided byWelchAllyn Noother use reproduction ordistributionofthispublication oranypartofit ispermittedwithoutwrittenpermission fromWelchAllyn Fortheaddresses andphonenumbersofallWelchAllynregionallocations visit welchallyn com locations ...

Reviews: