About Test-Quality Grades
Another type of feedback is the test-quality grade, as described in the following table.
If
Print Quality Grades
is enabled in the settings, a grade appears on printed reports and
also displays on screen when you view the values or interpretation of a completed test (as
described under
To learn how to enable or disable this setting, see
“Reviewing the Spirometry Print
Test-Quality
Grade
Number of Acceptable Efforts
Reproducibility
A
2 or more
Largest two FEV1 values match
≤
100 mL.
Largest two FVC values match
≤
100 mL.
B
2 or more
Largest two FEV1 values match > 100 and
≤
150 mL.
C
2 or more
Largest two FEV1 values match > 150 and
≤
200 mL.
D
1 or more
Largest two FEV1 values match > 200 mL.
F
None