Directions for Use
Chapter 4 Performing Spirometry Tests
41
6.
Press
Effort Type
as needed to select the type of effort you want the patient to
perform. See
•
FVC
•
FVC Post*
•
SVC
•
SVC Post*
*FVC Post and SVC Post are available only if you have already accepted at least one
pre-effort of the same type.)
7.
(FVC testing only) Press
Curve
as needed to select the curve type that you want to
view while testing. See
.
•
Flow/Volume
•
Volume/Time
•
Tidal Volume
•
Incentive
Figure 28. Spirometry Screen, Ready to Start Effort
8.
When ready, press
Start
.
9.
Coach the patient through the effort. For tips, see
10. When finished, you can press
Stop
, but you do not have to. The device stops
automatically when air stops moving (that is, when the ATS end-of-test criteria
are met).
Note
To learn how to change the default FVC curve type and parameters, see
“Reviewing the Spirometry Screen Settings”
FVC #1
9:17AM Oct 16 08
Doe, Jane
Curve
Flow/Vol
SVC #1
9:17AM Oct 16 08
Doe, Jane
NHANES III 1999
ATS: --
Pred
Value %Pred
FVC
2.09
--
--
FEV1
3.06
--
--
FEV1%
80.86
--
--
FEV6
3.75
--
--
PEF
7.14
--
--
FEF25-75 3.00
--
--
FEV0.5
2.31
--
--
FEV2
--
--
--
Value
SVC
--
ERV
--
IRV
--
VT
--
BF
--
Tin/Tex
--
Start
Effort Type
SVC
Effort Type
FVC
SVC example
FVC example (flow/vol curve)
Test
Done
Test
Done
Start
Vol (L)
Time (s)
5
5
Select the desired effort type.
Select the desired effort type and curve.
Flow (L/s)
Vol (L)
5
Summary of Contents for CP 200
Page 1: ...CP 200 Spirometry Option Directions for Use ...
Page 34: ...30 Chapter 3 Calibrating the Spirometer Welch Allyn CP 200 Spirometry Option ...
Page 52: ...48 Chapter 4 Performing Spirometry Tests Welch Allyn CP 200 Spirometry Option ...
Page 53: ...49 5 Troubleshooting Problem Solving Suggestions 50 Limited Warranty 51 Service Policy 51 ...
Page 56: ...52 Chapter 5 Troubleshooting Welch Allyn CP 200 Spirometry Option ...
Page 80: ...76 Appendix D Predictive Norms etc Welch Allyn CP 200 Spirometry Option ...
Page 91: ......