Chapter 3 Calibrating the Spirometer
Welch Allyn CP 200 Spirometry Option
4.
Connect a new flow transducer to the pressure tubing. See
5.
Attach the flow transducer to the syringe’s port, shown here. Push the flow
transducer all the way in for a tight seal.
Figure 12. Calibration Syringe
6.
Pull the plunger all the way out.
7.
Press
Continue
.
8.
Press
Start
.
9.
When the blue bar begins to move, push the plunger all the way in, then pull it all the
way out, carefully following the bar’s rate. Use a steady motion in both directions. See
Figure 13. Simulated Exhalation and Inhalation
Caution
Several things may affect calibration results: movement of the
syringe, movement of the pressure tubing, or blockage of air. Place the
syringe on a hard, level surface with at least 1 cubic meter of open air
surrounding the flow transducer. Place your hand on top of the syringe to
prevent movement.
Port
Plunger
Spirometer Calibration
9:17AM Oct 16 08
Target Rate
Stop
Push plunger in following the bar
Spirometer Calibration
9:17AM Oct 16 08
Target Rate
Stop
Pull plunger out following the bar
Summary of Contents for CP 200
Page 1: ...CP 200 Spirometry Option Directions for Use ...
Page 34: ...30 Chapter 3 Calibrating the Spirometer Welch Allyn CP 200 Spirometry Option ...
Page 52: ...48 Chapter 4 Performing Spirometry Tests Welch Allyn CP 200 Spirometry Option ...
Page 53: ...49 5 Troubleshooting Problem Solving Suggestions 50 Limited Warranty 51 Service Policy 51 ...
Page 56: ...52 Chapter 5 Troubleshooting Welch Allyn CP 200 Spirometry Option ...
Page 80: ...76 Appendix D Predictive Norms etc Welch Allyn CP 200 Spirometry Option ...
Page 91: ......