Directions for Use
Chapter 2 Reviewing the Spirometry Settings
15
Reviewing the Calibration Settings
To review or change the settings that affect calibration — or to calibrate the spirometer —
follow these steps.
1.
Press the Menu key
.
2.
Choose
Spirometry Settings
>
Calibration Settings
.
The following screen appears.
Figure 3. “Spirometry Calibration Settings” Screen
3.
Change any desired settings.
Setting
Description
Calibrate Spirometer
Brings up the
Spirometer Calibration
screen. See
Enable Auto Calibration Report
Yes or no. If yes, a calibration report prints automatically every time you accept
calibration results.
Print Calibration Report
Prints the most recent calibration report.
Spirometry Calibration Settings
9:17AM Oct 16 08
1 Calibrate Spirometer
2 Enable Auto Calibration Report
3 Print Calibration Report
0 Previous Menu