CP 150 spirometry option symptoms and solutions
Problem-solving suggestions:
If you try these suggestions and still have problems, contact Welch Allyn.
Note
Your model might not contain this feature.
Symptom
Possible cause
Suggested action
Unable to calibrate
Poor connection between flow
transducer and sensor
Check the connection between flow
transducer and sensor.
Damage to flow transducer
Replace the flow transducer if it is
damaged.
Leak during calibration.
Ensure that the connection between
the calibration syringe and flow
transducer is tight with no leaks.
Uneven calibration strokes.
Use even strokes in calibration.
Pressure tubing is kinked
Replace pressure tubing.
Calibration fails consistently. Unable to
achieve a ± 3% error percentage
margin
(with confirmed good syringe and
tubing and correct calibration
techniques).
Altitude and atmospheric pressure
beyond tolerance causes the
calibration reading to be above the 3%
error margin.
After a failed calibration, touch
Accept
even though the calibration
result is above the 3% error margin.
Accepting the calibration result will
adjust the gain value closer to the
correct atmospheric pressure. Perform
the calibration again.
No sensor detected
Poor connection between the sensor
and the device
Connect to another USB port.
Replace the USB cable.
Does not print
Out of paper
Load paper. See the CP 150
electrocardiograph manual, entitled
CP
150 12-lead resting electrocardiograph
Directions for use
.
Paper jam
If the paper is jammed, clear it, then
reload.
Values are too high (intermittent)
Patient’s fingers obstructed the screen
on the back of the flow transducer,
causing high back pressure and false
reading
Retest.
Patient’s lips were not tightly sealed
around the flow transducer
Retest.
Spirometer was calibrated with the
wrong size syringe
Recalibrate with a 3-liter syringe. See
Perform a calibration
in the spirometry
manual entitled
CP 150 spirometry
option, Directions for use
(English-
only).
Service manual
Troubleshooting 67
Summary of Contents for CP 150
Page 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Page 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Page 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Page 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Page 27: ...Home screen Home screen The Home screen includes the following areas 23...
Page 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Page 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Page 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Page 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Page 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Page 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Page 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Page 133: ......