52
Disassembly and Repair Procedures
Welch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Paper Tray Cover
Procedure
1.
Open the unit using the
directions.
2.
Next, remove the paper tray cover. See
Replacing the Printer Assembly (page 50)
3.
Slide the paper tray cover out until it stops.
See
.
Figure 37. Sliding the Paper Tray Cover
4.
Turn paper tray assembly up-side-down.
5.
Remove the 3 screws holding the paper
guide. See
.
Figure 38. Bottom of the Paper Tray Assembly
QuickStep
OperatingInstructions
StatECG
StatECG
Holddown
for1
second
AutomaticECG
Select
EnterPatient
Data
Select
DONE
Verify
ECGquality
Select
RhythmECG
Select
toend
Select
Changesettings
asdesired
ECG
IEC
C1
C2
C3
C4
C5
C6
R
L
N
F
AAMI
V1
V2
V3
V4
V5
V6
RA
LA
RL
LL
VI
V2
V3
R4
V4
V5
V6
LA
LL
RL
Paper Tray Cover
Screws holding paper guide
Bottom of paper tray cover
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......