Service Manual
Functional Verification
31
Battery Accuracy Test
When a voltage of 6.2 ± 0.05 VDC is supplied to the unit battery terminals, and the AC is
disconnected from the unit, the unit reports a “Normal to High” voltage.
When a voltage of 5.45 ± 0.05VDC is supplied to the unit battery terminals, and the AC is
disconnected from the unit, the unit reports a “critically low” voltage.
1.
Unplug unit from line power.
2.
Connect a DC power supply in place of the battery.
3.
Adjust power supply to 6.2VDC.
4.
Turn the unit on, and then verify the unit does not indicate a low or critically low
battery.
5.
Adjust power supply to 5.45 VDC.
6.
Verify the unit indicates a critically low power and will shut down.
Sleep/Standby Mode Current Test
When the unit is powered on and the AC is disconnected, and the unit is in low power
standby mode, the unit’s current is:
1.
Unplug the unit from line power and adjust the power supply to 6.4V DC ± .05.
2.
Connect the DVM in series with the positive battery connection from the power
supply.
3.
Press the on/standby button.
4.
Verify the LCD turns off and that the current draw is less than indicated for the unit.
5.
Disconnect the power supply.
Charger Detect Test
1.
Connect a DVM to the battery harness while observing correct polarity.
2.
Plug the unit into the AC power, and power ON.
3.
Choose
Main Menu
>
Systems Setting
>
Device info
>
Service Info
>
Service
Tests>PPPTEST2.EXE.
See
.
Note
The unit may have a 5 to 10 second delayed response.
CP 100 Current
CP 200 Current
< 5 mA
< 5 mA
Note
There may be up to a 10 sec delay in the drop of current due to the progressive
hibernation mode.
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......