Note
Disinfect according to your facility's protocols and standards or local regulations.
CaviWipes or Sani-Cloth Plus
1. Using CaviWipes or Sani-Cloth Plus, wipe the surface of the monitor to remove all debris.
2. Allow the monitor surface to dry for a minimum of 10 minutes before using the monitor.
70 percent isopropyl alcohol
Wipe the monitor with a clean cloth slightly dampened with 70 percent isopropyl alcohol.
10 percent chlorine bleach solution
1. Wipe the monitor with a clean cloth slightly dampened with a 10 percent bleach and water
solution. Follow the cleaning agent manufacturer's guidelines.
2. Rinse with a clean cloth slightly dampened with water that meets EP and USP quality
standards.
3. Allow the monitor surface to dry for a minimum of 10 minutes before using the monitor.
Clean the accessories
1. Wipe the NIBP hose and any reusable cuffs with a damp cloth moistened in a mild detergent
solution.
2. Wipe the temperature probe with a cloth dampened with alcohol, warm water, or an
appropriately diluted, nonstaining disinfectant solution.
3. Clean the SpO2 sensor with a cloth dampened with 70 percent isopropyl alcohol or 10 percent
chlorine bleach solution.
4. Clean the ear thermometer according to the manufacturer's directions for use.
5. Clean the ear thermometer dock with the same cleaning agents used on the monitor.
Caution
Never immerse any monitor accessories, such as NIBP hoses or cuffs,
temperature probes, or SpO2 sensors.
Clean the stand
The same cleaning agents used on the monitor can be used on the stand and accessory cable
management system. Clean on a routine basis according to your facility's protocols and
standards, or local regulations.
Note
Disinfect according to your facility's protocols and standards, or
local regulations.
Directions for use
Maintenance and service
91
Summary of Contents for Connex Vital Signs Monitor 6000 Series
Page 1: ...Welch Allyn Vital Signs Monitor 6000 Series Directions for use ...
Page 6: ...vi Contents Welch Allyn Vital Signs Monitor 6000 Series ...
Page 8: ...2 Introduction Welch Allyn Vital Signs Monitor 6000 Series ...
Page 14: ...8 Screen elements Welch Allyn Vital Signs Monitor 6000 Series ...
Page 22: ...16 Controls indicators and connectors Welch Allyn Vital Signs Monitor 6000 Series ...
Page 30: ...24 Setup Welch Allyn Vital Signs Monitor 6000 Series ...
Page 40: ...34 Navigation Welch Allyn Vital Signs Monitor 6000 Series ...
Page 44: ...38 Profiles Welch Allyn Vital Signs Monitor 6000 Series ...
Page 64: ...58 Alarms Welch Allyn Vital Signs Monitor 6000 Series ...
Page 92: ...86 Patient monitoring Welch Allyn Vital Signs Monitor 6000 Series ...
Page 98: ...92 Maintenance and service Welch Allyn Vital Signs Monitor 6000 Series ...
Page 106: ...100 Specifications Welch Allyn Vital Signs Monitor 6000 Series ...
Page 134: ...128 Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series ...
Page 143: ......