Battery power manager messages
Message
Possible cause
Suggested action
Low battery 5 minutes or less
remaining. (High-priority alarm)
Battery power is extremely low
Connect the monitor to AC
power. (If not connected to
AC power, the monitor powers
down when AC power is
depleted.)
Low battery 30 minutes or less
remaining.
Battery power is low
Touch
OK
to dismiss or
connect the monitor to AC
power.
Powering down. Call for
service.
Power manager or battery faults have
occurred
Call for service.
Battery is absent or faulty.
There is no battery in the monitor
Insert a battery.
The battery is faulty
Replace the battery.
Device is operating in battery
mode.
The AC power cord has been disconnected Touch
OK
to dismiss or
connect the monitor to AC
power.
Configuration Manager messages
Message
Possible cause
Suggested action
Unable to load configuration;
using factory defaults.
A configuration load error occurred
Call for service.
Functional error. Call for
service.
A critical configuration load error occurred Call for service.
No connection for send.
The monitor is not configured to the network Call for service.
Printer messages
Message
Possible cause
Suggested action
Low battery; plug into outlet.
The monitor's battery voltage is too low to
support printing
Connect the monitor to AC
power.
Printer door is open; close to
continue
The printer door is open
Close the printer door.
Out of paper.
The paper is not properly loaded
Align the paper with the print
head. If the problem persists,
call for service.
The paper sensor does not detect paper
Replace the paper. If the
problem persists, call for
service.
126
Troubleshooting
Welch Allyn Vital Signs Monitor 6000 Series™
Summary of Contents for Connex Vital Signs Monitor 6000 Series
Page 1: ...Welch Allyn Vital Signs Monitor 6000 Series Directions for use ...
Page 6: ...vi Contents Welch Allyn Vital Signs Monitor 6000 Series ...
Page 8: ...2 Introduction Welch Allyn Vital Signs Monitor 6000 Series ...
Page 14: ...8 Screen elements Welch Allyn Vital Signs Monitor 6000 Series ...
Page 22: ...16 Controls indicators and connectors Welch Allyn Vital Signs Monitor 6000 Series ...
Page 30: ...24 Setup Welch Allyn Vital Signs Monitor 6000 Series ...
Page 40: ...34 Navigation Welch Allyn Vital Signs Monitor 6000 Series ...
Page 44: ...38 Profiles Welch Allyn Vital Signs Monitor 6000 Series ...
Page 64: ...58 Alarms Welch Allyn Vital Signs Monitor 6000 Series ...
Page 92: ...86 Patient monitoring Welch Allyn Vital Signs Monitor 6000 Series ...
Page 98: ...92 Maintenance and service Welch Allyn Vital Signs Monitor 6000 Series ...
Page 106: ...100 Specifications Welch Allyn Vital Signs Monitor 6000 Series ...
Page 134: ...128 Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series ...
Page 143: ......