Ethernet messages
Message
Possible cause
Suggested action
Network not found; check
network cable connection.
A network cable is unplugged
Check the network cable
connection. If problem
persists, call for service.
A network connection is broken elsewhere
USB messages
Message
Possible cause
Suggested action
USB Communication failure.
Call for service.
An internal or external device is connected
but failed enumeration
Call for service.
External device not licensed for
use.
A license for an external device (e.g.,
barcode scanner) has not been activated
Disconnect the unlicensed
device.
External device not recognized. An unrecognized external device is
connected
Disconnect the unrecognized
device.
Incompatible Welch Allyn
device.
A communication protocol failure has
occurred
Call for service.
System messages
Message
Possible cause
Suggested action
Set date and time.
The date or time is not set
Set the date and time.
The date or time is not set properly
Reset the date or time.
Ambient temperature outside
operating range. Retry
measurement.
The ambient temperature is out of range
Operate the monitor within the
specified temperature range.
Retry patient temperature
measurement. If the message
persists, move the patient and
the monitor to a cooler
location.
Device shutdown is not
available at this time.
Device cannot perform an immediate
shutdown
Touch
OK
, wait, and retry.
Advanced settings unavailable Sensors are taking measurements
Stop continuous
measurements.
A physiological alarm condition is active
Respond to or reset the alarm.
Spot Check measurements have not been
saved
Save the measurements.
Directions for use
Troubleshooting
125
Summary of Contents for Connex Vital Signs Monitor 6000 Series
Page 1: ...Welch Allyn Vital Signs Monitor 6000 Series Directions for use ...
Page 6: ...vi Contents Welch Allyn Vital Signs Monitor 6000 Series ...
Page 8: ...2 Introduction Welch Allyn Vital Signs Monitor 6000 Series ...
Page 14: ...8 Screen elements Welch Allyn Vital Signs Monitor 6000 Series ...
Page 22: ...16 Controls indicators and connectors Welch Allyn Vital Signs Monitor 6000 Series ...
Page 30: ...24 Setup Welch Allyn Vital Signs Monitor 6000 Series ...
Page 40: ...34 Navigation Welch Allyn Vital Signs Monitor 6000 Series ...
Page 44: ...38 Profiles Welch Allyn Vital Signs Monitor 6000 Series ...
Page 64: ...58 Alarms Welch Allyn Vital Signs Monitor 6000 Series ...
Page 92: ...86 Patient monitoring Welch Allyn Vital Signs Monitor 6000 Series ...
Page 98: ...92 Maintenance and service Welch Allyn Vital Signs Monitor 6000 Series ...
Page 106: ...100 Specifications Welch Allyn Vital Signs Monitor 6000 Series ...
Page 134: ...128 Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series ...
Page 143: ......