1. Verify that you are using the Intervals Monitoring or Continuous Monitoring profile.
2. Touch the
Alarms
tab.
3. Touch the
Pulse rate
vertical tab.
4. Ensure that the Pulse rate alarm limit control is set to ON.
Note
If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.
5. Enter the desired upper and lower alarm limits for pulse rate using the up/down
arrow keys or the keypad.
6. Touch the
Home
tab.
The new alarm settings take effect immediately.
Manual parameters frame
The Manual parameters frame, located in the lower right of the Home tab, supports
manual entry of parameters and displays measurements taken by some accessories.
Note
Body mass index (BMI) is only available with an attached weight scale that
calculates BMI.
Note
When a measurement is transferred from an attached weight scale to the
device, the measurement displayed on the device is within one decimal
place (0.1) of the measurement displayed by the weight scale.
Note
You cannot manually enter temperature in this frame on a device
configured with a SureTemp Plus temperature module.
Note
You cannot manually enter respiration rate in this frame on a device
configured with CO2, RRa, or EarlySense.
Enter manual parameters
Note
The Manual parameters frame enables you to enter measurements taken
manually and displays measurements taken by some accessories. You can
select and configure the parameters in Advanced settings. Only four
parameters appear in the Manual parameters frame.
CAUTION
Weight scales attached to this monitor must be running on
battery power (battery type is specified in the weight scale manufacturer's
directions for use). Do not use the weight scale's external power supply.
CAUTION
Before weighing a patient on any weight scale connected to the
monitor, disconnect all sensors from the patient. Doing so ensures an
accurate weight measurement and significantly reduces electrostatic
discharges that might disrupt the monitor.
1. From the Home tab, touch anywhere within the Manual parameters frame.
Directions for use
Patient monitoring 175
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......