No breath detected alarm delay. Select the duration from the list box. This alarm
delay specifies the minimum duration of a no-breath condition to activate an
alarm signal. You can set the default in Advanced settings.
6. Touch the
Home
tab.
The new alarm settings take effect immediately.
IPI
The Integrated Pulmonary Index (IPI) integrates four parameters and their interactions to
provide a numeric value indicating a patient's overall respiratory status. The integrated
parameters are etCO2, RR, SpO2, and PR. The additional status information provided by
IPI enables you to intervene before values of individual parameters reach levels of clinical
concern.
Note
IPI is available for adult patients as well as three groups of pediatric
patients (1–3 years, 3–6 years, and 6–12 years), but it is not available for
neonatal/infant patients.
Note
Pediatric patient subtypes appear on the Patient Summary tab and on the
Home tab as a label in the IPI frame, but they do not appear in the Patient
frame.
Note
IPI is available only with the Continuous Monitoring profile.
The range of IPI is 1–10. These values can be interpreted as indicated in the following
table.
Index range
Patient status
10
Normal
8–9
Within normal range
7
Close to normal range; requires attention
5–6
Requires attention and may require intervention
3–4
Requires intervention
1–2
Requires immediate intervention
Source:
The microMediCO2 Module - OEM Manufacturers' Integration Guide
(p. 13), Oridion Medical 1987 Ltd.
Note
The interpretation of a patient's IPI value might change in different clinical
conditions. For example, patients who have a history of respiratory
difficulties might require a lower IPI Low Alert threshold that takes into
account their diminished respiratory capacity.
IPI frame
The IPI frame displays a numeric value from 1–10 as well as an alarm threshold and
control.
128 Patient monitoring
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......