33 - Troubleshooting
Error codes
Error description of the ABPM 7100
Error symptom
Possible cause
Remedy
Time and date are not
updated following a longer
period without power
supply from power packs
or batteries.
The internal buffer battery is
depleted.
Date and time can be reset after every power
pack or battery replacement.
Send the device to your Welch Allyn specialist.
Measurement data can no
longer be called
up/displayed.
An error occurred during patient
data storage.
Delete the respective patient (menu bar) and
recreate it.
The connection between
the ABPM 7100 and the
PC is faulty.
The incorrect COM interface is set.
Set the correct interface in the service
programs.
Cable plug or socket is defective.
Inspect the plug and the socket on the ABPM
7100. Ensure that the pins are straight to
guarantee contact.
The ABPM 7100 is not in
transmitting mode (the displays
shows the time).
Switch the ABPM 7100 off and then on again
without removing the connecting cable.
No patient number.
The ABP Monitor is not initialized,
i.e. the patient number was not
transferred during the preparation
for a 24-hour measurement.
The patient number can also be transmitted
after the measurement. This does not
influence the measurement data.
No measurements were
conducted during the
nocturnal phase.
The battery packs or batteries
were prematurely depleted.
The power packs or batteries may be defective
(please contact your Welch Allyn specialist).
The patient has switched off the
ABPM 7100.
Draw the patient’s attention to the urgency of a
complete 24-hour measurement.
The display does not show
“
co
”.
You are not in transmitting mode.
Communication via cable: Switch the ABPM
7100 off and then on again without pulling the
plug.
No automatic
measurements will be
performed.
No manual measurements
performed after application.
Valid manual measurement must always be
performed after the device has been
positioned.
Incorrect log set.
Set log 1 or 2.
The measurement interval
does not meet your
expectations.
Incorrect log set.
The programmed log does not correspond with
the set log in the ABPM 7100. Check the log
manually on the device.
No manual measurements
performed after application.
Conduct manual measurement to activate the
set log
Troubleshooting - 34
Error symptom
Possible cause
Remedy
Err 1
The patient displays severe arrhythmia.
ABP Monitor not applicable.
The arm was moved during
measurement.
Keep the arm still during measurement.
Insufficient valid pulse rate detected.
Place the cuff on your arm again.
Err 2
The arm was moved during
measurement.
Keep the arm still during measurement.
Cuff does not fit the arm snugly.
Check the seating of the cuff and that of the
device.
Err 3
Blood pressure beyond the
measurement range.
Permanent notification renders the ABP Monitor
unsuitable for the patient.
Strong arm movements.
Keep the arm still during measurement.
Problems with the pneumatics.
If the error persists permanently, send the device
to your Welch Allyn specialist.
Err 4
Data transmission cable incorrectly
inserted in ABP Monitor.
Insert the cable into the ABP Monitor correctly.
Pins in the plug of the data transmission
cable are mechanically damaged.
Check the plug to see whether the pins on the
inside are damaged. If they are, contact your
Welch Allyn specialist.
Measurement value was not correctly
transmitted.
Restart the transmission.
Err 5
bAtt
Power pack or battery voltage too low.
Replace the power packs or batteries.
Power packs or batteries are defective.
The power pack or battery voltage is correct but
“
bAtt
” is displayed during cuff inflation. Replace
the power packs.
Battery contacts are corroded.
Clean the battery contacts with a cotton cloth and
a little alcohol.
Err 6
+
Possible continuous
alarm until a button is
pressed.
Build-up 34 fair.
Check the cuff for a build-up of air or a kink in the
tubing. If the cuff tubing is kinked, straighten the
tubing. Otherwise send the device in for
inspection immediately.
Blood pressure cuff incorrectly
connected.
Connect the cuff to the device.
Leaky points in the cuff or the cuff
tubing.
If necessary, replace the cuff.
Err 7
The memory of the blood pressure
measurement device is full (a maximum
of 300 measurements and events can
be saved; with CBP or PWA: 260
measurements).
Delete the data in the ABP Monitor but ensure
that the data was stored on your PC first.
Err 8
Measurement cancelled with a pressed
button.
Err 9
+
Possible continuous
alarm until a button is
pressed.
Residual pressure inside the cuff
Wait for the cuff to deflate completely.
Zero point comparison was
unsuccessful.
Send the device to your specialist for inspection
immediately or directly to your Welch Allyn
specialist.
EN
Summary of Contents for ABPM 7100
Page 2: ......
Page 26: ...Material No DIR 722592 80019691 Ver E Revision date 2022 02...
Page 50: ...Material No DIR 722618 80019699 Ver E Revideret 2022 02...
Page 74: ...Material No DIR 722614 80019695 Ver E revisiedatum 2022 02...
Page 98: ...Material No DIR 722621 80019702 Ver E p ivitetty 2022 02...
Page 122: ...Material No DIR 722619 80019700 Ver E Date de r vision 2022 02...
Page 146: ...Material No DIR 722611 80019692 Ver E Revisionsdatum 2020 02...
Page 170: ...Material No DIR 722620 80019701 Ver E Data revisione 2022 02...
Page 194: ...Material No DIR 722615 80019696 Ver E Revisjonsdato 2022 02...
Page 218: ...Material No DIR 722617 80019698 Ver E Data de revis o 2022 02...
Page 242: ...Material No DIR 722616 80019697 Ver E Fecha de revisi n 2022 02...
Page 266: ...Material No DIR 722980 80019694 Ver E revisionsdatum 2022 02...
Page 290: ...Material No DIR 409846 80021803 Ver E Revizyon Tarihi 2022 02...
Page 314: ...Material No DIR 774666 80029425 Ver E datum revize 2022 02...
Page 322: ...15 ABPM 7100 ABPM 7100 MRI ABPM 7100 ABPM 7100 16 30 C 45 C 0 C 24 2 6 V NiMH 3 10 V...
Page 327: ...25 ME ME 26 30 C ABPM 7100 MRI ABPM 7100 ABPM 7100 HF EL...
Page 335: ...ABPM 7100 ABPM 7100 A V ME ME ABPM 7100 ABPM 7100 MRI ABPM 7100 ABPM 7100 EL...
Page 338: ...Material No DIR 774668 80029426 E d 2022 02...
Page 362: ...Material No DIR 774668 80029427 Ver E datum verzije 2022 02...
Page 386: ...Material No DIR 774668 80029428 E verzi Verzi fel lvizsg lati d tuma 2022 02...
Page 394: ...15 ABPM 7100 ABPM 7100 ABPM 7100 ABPM 7100 16 30 C 45 C 0 C 24 NiMH 2 6 3 10...
Page 399: ...25 ME ME 26 30 C ABPM 7100 ABPM 7100 ABPM 7100 KK...
Page 407: ...41 ABPM 7100 ABPM 7100 A V ME ME ABPM 7100 42 ABPM 7100 ABPM 7100 ABPM 7100 KK...
Page 410: ...Material No DIR 774691 80029429 E 2022 02...
Page 434: ...Material No DIR 774692 80029430 Vers E p rskat anas datums 2022 02...
Page 458: ...Material No DIR 774693 80029431 wer E data aktualizacji 2022 02...
Page 466: ...15 ABPM 7100 ABPM 7100 27B ABPM 7100 ABPM 7100 16 30 C 28B 45 C 0 C 24 2 6 3 10...
Page 471: ...25 38B 26 30 C ABPM 7100 ABPM 7100 ABPM 7100 RU...
Page 479: ...41 ABPM 7100 ABPM 7100 ABPM 7100 42 ABPM 7100 ABPM 7100 ABPM 7100 RU...
Page 482: ...Material No DIR 774694 80029432 E 2020...
Page 483: ......
Page 484: ...Material No DIR 722980 80019703 Ver E Revision Date 2022 02...