CONFORMANCE TESTING
97
Battery Performance Information
With a new, fully charged lithium-ion battery, the ELI 380 is typically capable of acquiring more than
30 resting ECGs with 1 performed every 10 minutes before a recharge is necessary. When two lithium-ion
batteries are used, more than 60 resting ECGs may be acquired with 1 performed every 10 minutes before a
recharge is necessary.
For optimal performance, connect the ELI 380 to AC power whenever it is not in use. The device can be
used with AC power while simultaneously recharging the battery/batteries.
Functional Testing
MAIN SCREEN:
The MAIN SCREEN is displayed when the unit is first turned on.
The LCD will timeout and go dark if there is no ECG or user input. Touch the touchpad to re-activate.
SERVICE SCREEN:
To access the SERVICE SCREEN begin at the MAIN SCREEN.
Press
.
Press
ADVANCED,
then enter the Admin password.
Press
SERVICE
.
CONFIGURATION SCREEN:
To access the CONFIGURATION SCREEN begin at the MAIN SCREEN.
Press
.
Press
ADVANCED,
then enter Admin password.
Display and Standby Mode
Press the ON button and verify the text on display is clear and legible and there are no flickering or missing
lines/pixels. Record Results on the TDR
Close (tilt) the LCD Display towards the writer. After a few seconds verify the LCD backlight turns off.
Open the LCD Display and verify the unit comes out of the standby mode returning to the MAIN
SCREEN. Record Results on the TDR
The following procedure applies to the ERGO variant:
1.
Rotate the screen 90 degrees away from the front user face to the right and verify that the LCD
backlight is on with no flickering or missing lines/pixels. Record Results on the TDR
2.
Rotate the screen 90 degrees away from the front user face to the left and verify that the LCD
backlight is on with no flickering or missing lines/pixels. Record the Results on the TDR
3.
Tilt the display out until a mechanical stop is reached. Verify the stop is reached at 120 degrees.
Record Results on the TDR
Auto Test
From the Service Screen, select
Auto Test
. Verify the auto test completes one cycle without an error.
The Auto Test function tests the ELI 380’s ability to read and write an ECG file to the flash memory in the
unit. Record Results on the TDR