38
Welch Allyn Spot Vital Signs
Table 3. Temperature
Code
Description
Corrective Action
C20
Broken/missing probe
Replace probe.
P
Loss of tissue contact
Ensure proper probe positioning.
E0.2, E0.3
Ambient temperature out of range
Adjust ambient temperature or device location.
C22
10-minute diagnostic time exceeded
Remove probe, discard probe cover, retake
temperature.
Table 4. SpO
2
Code
Description
Corrective Action
E7
Internal SpO
2
error.
Retake reading.
C6
SpO
2
pulse rate out of range
Check patient condition.
C8
Faulty SpO
2
sensor.
Replace sensor.
C9
SpO
2
time limit exceeded.
Remove sensor from patient. Reapply sensor and
retake reading.