44
Welch Allyn Spot Vital Signs
Pulse Oximetry
Masimo Sensor Accuracy Guide
Accuracy specified when used with Masimo SET pulse oximetry monitors or with
licensed Masimo SET pulse oximetry modules using Masimo patient cables, during no
motion. Numbers present ± 1 standard deviation. Plus or minus one standard deviation
represents 68% of the population. SpO
2
accuracy from 70% to 100%. Pulse rate
accuracy from 25 to 240 bpm.
Masimo Patents
The Masimo sensors and cables are covered under one or more of the following U.S.A.
patents: 5,758,644; 5,823,950; 6,011,986; 6,157,850; 6,263,222; 6,501,975; and other
applicable patents listed at www.masimo.com/patents.htm.
Performance Measurement Range
SpO
2
: 70 to 100%
Pulse Rate: 25 - 240 beats per minute (BPM)
Perfusion
0.02% to 20%
SpO
2
Accuracy
Saturation: 70% to 100%
No Motion: Adults, Pediatrics ± 2 digits
Motion: Adults, Pediatrics ± 3 digits
Low Perfusion: Adults, Pediatrics ± 2 digits
Pulse Rate Accuracy
Pulse Rate: 25 to 240 bpm
No Motion: Adults and Pediatrics ± 3 digits
Motion: Adults and Pediatrics ± 5 digits
Low Perfusion: Adults and Pediatrics ± 3 digits
Table 14. Masimo Sensor Accuracy Guide
Saturation Accuracy
Pulse Rate Accuracy
Sensor
Weight Range
No Motion
Motion
No Motion
Motion
LNCS-DCI
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
LNCS-DCIP
10 to 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
LNCS-ADTX
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
LNCS-PDTX
10 to 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
LNCS INF-L
3 to 20 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
LNOP-DCI
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
LNOP-DCIP
10 to 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
LNOP-ADT
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
LNOP-PDT
10 to 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
LNOP INF-L
3 to 20 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
Summary of Contents for 4200B
Page 3: ...Welch Allyn Spot Vital Signs Directions for Use...
Page 6: ...iv Welch Allyn Spot Vital Signs...
Page 22: ...14 Welch Allyn Spot Vital Signs...
Page 28: ...20 Internal Configuration Welch Allyn Spot Vital Signs...
Page 36: ...28 Welch Allyn Spot Vital Signs...
Page 42: ...34 Welch Allyn Spot Vital Signs...
Page 44: ...36 Welch Allyn Spot Vital Signs...
Page 50: ...42 Welch Allyn Spot Vital Signs...
Page 60: ...52 Welch Allyn Spot Vital Signs...
Page 70: ...62 Welch Allyn Spot Vital Signs...
Page 72: ...64 Warranty Welch Allyn Spot Vital Signs...
Page 73: ......