8
37
Error Indications and Interpretation
The following table of conditions and error codes provides a quick reference of the
descriptions and probable causes of error codes.
To clear the error code:
Power the Spot Vital Signs off, wait five seconds, and power on. If the error code
reappears then power the Spot Vital Signs off and disconnect the battery for five minutes.
Reconnect the battery and power on. If the error code continues to reappear, call Welch
Allyn for a Service Notification Number (see “Technical Assistance” on page 56).
Press the Blood Pressure Start/Stop button to reset flashing patient alarm conditions.
Error Codes
Table 1. General
Code
Description
Corrective Action
E11
Internal safety violation
Check patient, contact Technical Service.
C12
Ambient temperature out of range
Adjust ambient temperature or device location.
C13
Battery failure
Use wall transformer.
E0.0 - E9.9
Temperature module malfunction
Contact Technical Service.
E42
Internal communications error
Disconnect the battery and wait 5 minutes.
Reconnect the battery and then set the date and time,
see “Blood Pressure Hose and Cuff Connections” on
page 21.
E20 - E50
General internal malfunction
Contact Technical Service.
Table 2. Blood Pressure
Code
Description
Corrective Action
C02
Auto-zero failure
Check for air obstruction, limit patient movement.
C03
Inflation too rapid
Check for kinked blood pressure cuff tubing, pressure
hose, or other air obstruction.
C04
Excessive inflation time
Check for air leaks.
C05
Excessive noise
Check patient condition, blood pressure cuff
placement, limit patient movement.
C06
Measurement was outside of device’s
measurement range
Check patient condition.
E10
Blood pressure cuff overpressure condition
Check patient condition.
Summary of Contents for 4200B
Page 3: ...Welch Allyn Spot Vital Signs Directions for Use...
Page 6: ...iv Welch Allyn Spot Vital Signs...
Page 22: ...14 Welch Allyn Spot Vital Signs...
Page 28: ...20 Internal Configuration Welch Allyn Spot Vital Signs...
Page 36: ...28 Welch Allyn Spot Vital Signs...
Page 42: ...34 Welch Allyn Spot Vital Signs...
Page 44: ...36 Welch Allyn Spot Vital Signs...
Page 50: ...42 Welch Allyn Spot Vital Signs...
Page 60: ...52 Welch Allyn Spot Vital Signs...
Page 70: ...62 Welch Allyn Spot Vital Signs...
Page 72: ...64 Warranty Welch Allyn Spot Vital Signs...
Page 73: ......