WM
6807
1
0
5
/2
021
MEDUMAT Standard
2
EN
141
7 Description of the modes
There are the following trigger options in the individual ventilation
modes:
7.2 Ventilation parameters
Ventilation
mode
Inspiration trigger
Expiration
trigger
Trigger window
for mandatory
breaths
ASB breath
IPPV
No
No
No
No
B ASB
Yes
Yes
20 % of Te
Yes
CPAP
No
No
No
No
CPAP + ASB
Yes
Yes
No
Yes
aPCV
Yes
No
Can be set from
0 % - 100 % of Te
No
PCV
No
No
No
No
PRVC + ASB
Yes
Yes
20 % of Te
Yes
SIMV
Yes, permanently set
No
20 % of Te
No
SIMV + ASB
Yes
Yes
20 % of Te
Yes
S-IPPV
Yes (with Flow
measu ASB
option can be set)
No
100 % of Te
No
Demand
Yes, permanently set
Yes, permanently set No
No
CCSV
Yes
No
No
No
Ventilation
parameters
Unit
Description
Vt
ml
Tidal volume (breath volume)
Freq.
1/min
Ventilation rate
pMax
mbar
Maximum ventilation pressure
PEEP
mbar
Positive end-expiratory pressure (CPAP)
Air Mix
-
Ventilation through the addition of
ambient air
93 % oxygen
-
Ventilation with concentrator oxygen
100 % oxygen
-
Ventilation with 100 % oxygen
Δ
pASB (only with
Flow
measu ASB
option)
mbar
Pressure support (relative to the set PEEP)