background image

MANUVAC

Suction Pump

Description and instructions for use

Summary of Contents for MANUVAC

Page 1: ...MANUVAC Suction Pump Description and instructions for use...

Page 2: ...to operation 8 4 2 Aspiration 9 4 3 After aspiration 11 5 Hygienic preparation 12 6 Functional check 14 7 Troubleshooting 15 8 Maintenance 16 8 1 Intervals 16 8 2 Sending in device 16 8 3 Disposal 17...

Page 3: ...6 Suction valve 4 Connecting hose 10 Grip bar 8 Discharge valve 18 Vent flap 16 Thick suction hose length 280 mm 2nd stage 5 Holding clip for secretion bottle 13 Fine suction hose length 1020 mm 1st s...

Page 4: ...4 EN Overview 1 1 Special symbols on the device 1 No Symbol Meaning MANUVAC device information plate 1 SN Serial number of device Year of manufacture CE marking...

Page 5: ...risk of these sub stances being aspirated into the lungs In addition the MANUVAC vacuum pump can be used to evacuate vacuum mattresses MANUVAC is not intended to be used for bronchial aspiration 2 2...

Page 6: ...ety Regulations Directive 93 42 EEC we may point out the following Read these operating instructions carefully They belong to the appliance and must be kept avail able at all times Any use of the appl...

Page 7: ...uring the aspiration process take particular care not to cause injuries to the patient s mouth or throat e g to mucous membrane The suction can be interrupted by briefly open ing the fingertip for exa...

Page 8: ...he pedal and push the retaining clip outwards The pedal then springs up into the starting position 3 Lift the thick suction hose out of the curved housing groove 4 Aspiration is carried out by rhythmi...

Page 9: ...clinging to the skin You can leave the fingertip open all the time and keep your thumb over it To release the suction all you have to do is lift your thumb For effective aspiration of viscous and sol...

Page 10: ...ced or held upright secretion cap pointing upward If nessary the secretion bottle can also be held vertically or obliquely however the secretation cap must point upwards If the secretion cap is held d...

Page 11: ...ddle line of the secretion bottle then fit the hoses together Important Unless absolutely unavoidable the vacuum pump should never be transported with a full secretion bottle since the aspirated liqui...

Page 12: ...ient for the external surfaces Observe the instructions regarding use of disinfect ant For immersion disinfection we recommend gigasept FF new and terralin protect for wipe disinfection You are recomm...

Page 13: ...p In case of serious con tamination we recommend a precleaning with clear water Procedure is the same as for rinsing with disinfectant solution The pump does not need to be dismantled for this cleanin...

Page 14: ...e pump is disinfected and sterilised Procedure Ensure that the hose connections are tight Check that the hoses and rubber parts are in good condition Porous and brittle parts must be replaced Check th...

Page 15: ...the red vent flap of the secretion bottle fits tightly Check the disk diaphragm for cracks and replace if necessary see chapter 8 Maintenance on page 16 Check the hose system and rubber parts for crac...

Page 16: ...func tioning satisfactorily if no cracks or other defects can be seen Should it be found to be defective it can easily be changed by authorized specialist staff Have servicing and repair work perform...

Page 17: ...N Emergency Note If you send in obviously contaminated parts these will be disposed of by WEINMANN Emergency or by specialist staff expressly so authorized by WEINMANN Emergency at your expense 8 3 Di...

Page 18: ...wing accessories are not part of the stan dard product and must be ordered separately Set disposable aspiration catheters WM 15051 each consisting of 3 aspiration catheters Fr 06 2 mm Fr 10 3 mm Fr 12...

Page 19: ...mm WM 10801 Hose steramide diam 4 7 mm length 740 WM 5776 Hose silicone diam 10 15mm length 280 mm WM 2201 Adapter n d 10 WM 4627 Angled hose connection WM 2252 Secretion socket with fingertip control...

Page 20: ...e 160 ml stroke Max negative pressure 700 mbar Volume of secretion bottle 370 ml Temperature range Operation Storage 18 C to 50 C 40 C to 60 C Overflow protection device ball float Connecting hose len...

Page 21: ...uest In the event of a claim under warranty contact your specialist dealer 12 Declaration of Conformity WEINMANN Emergency Medical Technology GmbH Co KG declares herewith that the product com plies fu...

Page 22: ......

Page 23: ......

Page 24: ...for Production Logistics Service WEINMANN Emergency Medical Technology GmbH Co KG Siebenst cken 14 24558 Henstedt Ulzburg GERMANY Manufacturer WEINMANN Emergency Medical Technology GmbH Co KG Frohb se...

Reviews: