10
Checking the device
4.2 Preparations for testing
1. Boot up PC.
2. Start the PC software.
Overview of software versions:
•
EasyView
Reading, editing and saving usage data,
modifying device settings.
•
EasyView-Service
Works like EasyView, facility for programming the
serial number.
•
EasyView-Light
Works like EasyView; device settings cannot be
modified.
4.3 Entering the device data
Enter the serial number of the device (Device-No.) and
the manufacturing date into the test records.
4.4 Checking the devices self testing
1. Insert rechargeable or Battery-Pack.
2. Open the device lid.
Request:
– The device is giving off a short double peeping
tone.
– All LEDs light up briefly.
– The device announces clearly and completely
the spoken order “This device will assist you.”.
– The green status LED lights up.
3. Close the device lid.
EasyView
EasyView-Service
EasyView-Light
Summary of Contents for Homecare MEDUCORE Easy