Technical data
55
10. Technical data
VENTI
motion
/VENTI
logic
VENTI
motion
/VENTI
logic
with
VENTI
click
Product class as per EC directive
93/42/EEC
IIa
Dimensions WxHxD in cm
23 x 12.5 x 34
23 x 12.5 x 45.5
Weight
approx. 4.5 kg
approx. 4.8 kg
Temperature range
– operation
– storage
+5 °C to +35 °C
–40 °C to +70 °C
Air pressure range
750 – 1100 hPa
(corresponds to a height of approx. 2500 metres)
Max. respiration pressure at an air
pressure of 600 hPa
25 hPa
Electrical rating
115 – 230 V AC, 50–60 Hz
Tolerance -20 % +10 %
Power consumption during
– operation
– standby
230 V
115 V
0.2 A
0.4 A
0.043 A 0.068 A
230 V
115 V
0.35 A
0.7 A
0.043 A 0.068 A
Classification as per
EN 60601-1
– type of protection against
electric shock
– degree of protection against
electric shock
Protection class II
Type B
Electromagnetic
compatibility (EMC) as per
EN 60601-1-2:
– radio interference suppression
– radio interference immunity
EN 55011
EN 61000-3-2, EN 61000-3-3, EN 61000-4-2 to 6,
EN61000-4-8 EN61000-4-11
Mean sound pressure level/oper-
ation as per EN ISO 17510
at a distance of 1 m from the
device in the patient position
approx. 35 dB (A) at 20-35 hPa
approx. 31 dB (A) at 20 hPa
approx. 29 dB (A) at 15 hPa
approx. 27 dB (A) at 12 hPa
approx. 25 dB (A) at 10 hPa
approx. 23 dB (A) at 7 hPa
Sound pressure level, alarm
approx. 62 dB (A)
IPAP pressure range
EPAP pressure range
CPAP pressure range
Pressure accuracy
Increment
6 to 35 hPa
4 to 20 hPa
4 to 20 hPa
± 0.6 hPa
0.2 hPa
(1 hPa = 1 mbar
≈
1 cm H
2
O)
Minimum stable limit pressure
(PLS
min
) (min. pressure in the event
of a fault)
Maximum stable limit pressure
(PLS
max
) (max. pressure in the
event of a fault)
≥
0 hPa
≤
60 hPa
Respiratory frequency
Accuracy
Increment
6 to 45 l/min
± 0.5 l/min
1 l/min