2. Description
2.1 Purpose
With the “Re-usable Canister“ conversion set
you can convert the ACCUVAC
®
aspirator to
use re-usable canister instead of the dispos-
able canister.
2.2 Function
The re-usable collection canister is fixed to the
side of the motor unit and directly connected
to the suction port of the motor unit. There is
thus no need for an intermediate tube.
A replaceable hydrophobic bacteria filter
droplets of moisture from finding their way into
the motor unit and passing into the environ-
ment via the muffler.
The bacteria filter
use and sterilisation.
Important:
Do not immerse the bacteria filter
in disinfectant liquid, as this adversely
affects its hydrophobic properties.
An overfill system prevents secretions from
entering the motor unit. The ball
surface of the secretion until it blocks the exit.