6.2 Cleaning, disinfection and
sterilization
The individual parts may be cleaned,
disinfected and sterilized as shown in the
following table.
Observe the instructions regarding use of
disinfectant. We recommend GIGASEPT FF.
(1) Always dry the filter before using it again.
A damp filter reduces the pump’s suction
capacity.
(2) Do not use any tensides or alcohol for cleaning,
as they can adversely affect the hydrophobic
properties of the filter.
(3) After disinfection, rinse the parts thoroughly
with distilled water and let them dry.
(4) Hot steam sterilization at 134 °C with devices
in accordance with EN 285, stop time: 5
minutes.
Part
Cleaning
Disin-
fecting
Rinse in
washing
machine
Steriliza-
tion
Collection
canister
In warm
water
with mild
house-
hold clea-
ner
Immerse
in dilute
solution
(3)
Rinse at
up to
95 °C
Hot
steam ste-
rilization
up to
134 °C
(4)
Secretion
cover
Sealing ring
Overfill ball
Filter cover
Aspiration
tube
Filter (1)
In clear
warm
water (2)
Not permitted (2)
Nozzle with
fingertip
Disposable item, re-use not permitted. Use
new part for every patient.