Brine Nebulization System, Type Soldos V3
Index: 03
Change date: 01/03/2017
Operating instructions no.: BA DW 005-03 Soldos V3 EN.docx
Page 16 of 70
ATTENTION !
Do not kink the dosing line when routing it to the room! It is
essential to re-install a new dosing line when this happens during
the installation.
The length of the dosing line affects the pump operation time in the
disinfection program. It is therefore necessary to note the length of
hose installed in the disinfection protocol in
Section 9.5.1.
4.5
Electrical installation
DANGER DUE TO HIGH VOLTAGE !
The system must be switched to a voltage-free state before any
electrical work is performed!
ATTENTION !
The electronic components in the Soldos V3 controller are sensitive
to electrostatic discharge. The generally known precautions
regarding ESD-sensitive devices must be adhered to when handling
these electronic components.
This implies in particular:
Do not pull plug-in connections while they are live.
Discharge yourself for at least 5 seconds before you directly touch these
devices, e.g. by touching a grounded part of the system or by wearing an ESD
discharge band that is connected to ground.
Assembly:
The Soldos V3 is connected to the power mains via a safety power socket
(230V, 50Hz).
Electrical wiring: See circuit diagram and terminal plan in Section
9.2, Terminal
plans.
Connection of an external control signal for "Light and sound effects" according
to the terminal plan supplied.
Additional work for a connection to an external control centre (building
management system)
Connection of the floating collective fault message to the building management
system according to the terminal plan supplied.