![Waygate Technologies Krautkramer USM Go Operating Manual Download Page 167](http://html.mh-extra.com/html/waygate-technologies/krautkramer-usm-go/krautkramer-usm-go_operating-manual_878369167.webp)
Distance-amplitude correction (DAC)
5 Operation
USM Go
Rev. 10 (06/2021)
5-91
5.18 Distance-amplitude correction
(DAC)
The USM Go has an optional DAC function for the echo
evaluation by means of the distance-amplitude correc-
tion (DAC).
Due to the angle of beam spread and the sound attenu-
ation in the material, the echo height of equally sized re-
flectors depends on the distance to the probe.
A distance-amplitude correction curve, recorded using
defined reference reflectors, is a graphical representa-
tion of these influencing factors.
If you use a reference block having artificial flaws to re-
cord a DAC curve, you can use this echo amplitude for
the evaluation of a flaw without any further corrections.
The reference block should be made of the same mate-
rial as the test object.
The time-corrected gain TCG increases the gain in dis-
tance sensitive mode so that all reference echoes reach
80 % screen height. The amplitude evaluation of echo
indications is made in relation to the first reference echo.
– Switch to the second operating level.
– Select the function group
EVAL
.
– Select the function
EVAL MODE
and press a function
key to choose
DAC/TCG
.
– Switch back to the first operating level. The function
group
DAC/TCG
is now displayed there.
If a stored DAC curve exists, it becomes effective now.
Summary of Contents for Krautkramer USM Go
Page 1: ...Krautkrämer USM Go USM Go Operating Manual Id No 49 155 Rev 10 06 2021 ...
Page 5: ...USM Go Rev 10 06 2021 0 5 Second operating level ...
Page 6: ...0 6 Rev 10 06 2021 USM Go Second operating level continued ...
Page 10: ...0 10 Rev 10 06 2021 USM Go ...
Page 20: ...0 10 Rev 10 06 2021 USM Go ...
Page 21: ...USM Go Rev 10 06 2021 1 1 Introduction 1 ...
Page 35: ...USM Go Rev 10 06 2021 2 1 Standard package and accessories 2 ...
Page 41: ...USM Go Rev 10 06 2021 3 1 Initial start up 3 ...
Page 53: ...USM Go Rev 10 06 2021 4 1 Principles of operation 4 ...
Page 76: ...4 Principles of operation Saving the settings 4 24 Rev 10 06 2021 USM Go ...
Page 77: ...USM Go Rev 10 06 2021 5 1 Operation 5 ...
Page 210: ...5 Operation Evaluation according to the DGS method 5 134 Rev 10 06 2021 USM Go ...
Page 211: ...USM Go Rev 10 06 2021 6 1 Documentation 6 ...
Page 240: ...6 30 Rev 10 06 2021 USM Go ...
Page 241: ...USM Go Rev 10 06 2021 7 1 Maintenance and care 7 ...
Page 247: ...USM Go Rev 10 06 2021 8 1 Interfaces and Peripherals 8 ...
Page 251: ...USM Go Rev 10 06 2021 9 1 Appendix 9 ...
Page 266: ...9 16 Rev 10 06 2021 USM Go ...
Page 267: ...USM Go Rev 10 06 2021 10 1 Specifications 10 ...
Page 278: ...10 12 Rev 10 06 2021 USM Go ...