5.13 Defining the probe angle
You can use the function
AUTOANG
to define the cur-
rent index angle of a probe on a reference block. The
current probe index angle is influenced, e.g. by different
materials or by the wear of the probe contact face.
ATTENTION
You have to carry out calibration before us-
ing the function
AUTOANG
– After calibration, switch to the function group
AU-
TOANG
.
– Select the function
BLOCK
and press a function key
to choose the range of the calibration standard that
includes the nominal value of the probe angle (for ex-
ample K2 30-65 for the nominal value 45°).
– Shift the gate to the calibration echo.
– Switch to the function
RECORD
and press a function
key to record the calibration echo.
– Peak the echo. The USM Go will automatically record
the time-of-flight of the echo with the maximum ampli-
tude occurring during this.
– Press a function key to set the function
RECORD
back to
OFF
.
The calculated angle is briefly displayed at the bottom
edge of the display screen.
The values of the last measurement are automatically
entered in the function group
TRIG
, i.e. in the function
PROBE ANGLE
there.
Summary of Contents for Krautkramer USM Go
Page 1: ...Krautkrämer USM Go USM Go Operating Manual Id No 49 155 Rev 10 06 2021 ...
Page 5: ...USM Go Rev 10 06 2021 0 5 Second operating level ...
Page 6: ...0 6 Rev 10 06 2021 USM Go Second operating level continued ...
Page 10: ...0 10 Rev 10 06 2021 USM Go ...
Page 20: ...0 10 Rev 10 06 2021 USM Go ...
Page 21: ...USM Go Rev 10 06 2021 1 1 Introduction 1 ...
Page 35: ...USM Go Rev 10 06 2021 2 1 Standard package and accessories 2 ...
Page 41: ...USM Go Rev 10 06 2021 3 1 Initial start up 3 ...
Page 53: ...USM Go Rev 10 06 2021 4 1 Principles of operation 4 ...
Page 76: ...4 Principles of operation Saving the settings 4 24 Rev 10 06 2021 USM Go ...
Page 77: ...USM Go Rev 10 06 2021 5 1 Operation 5 ...
Page 210: ...5 Operation Evaluation according to the DGS method 5 134 Rev 10 06 2021 USM Go ...
Page 211: ...USM Go Rev 10 06 2021 6 1 Documentation 6 ...
Page 240: ...6 30 Rev 10 06 2021 USM Go ...
Page 241: ...USM Go Rev 10 06 2021 7 1 Maintenance and care 7 ...
Page 247: ...USM Go Rev 10 06 2021 8 1 Interfaces and Peripherals 8 ...
Page 251: ...USM Go Rev 10 06 2021 9 1 Appendix 9 ...
Page 266: ...9 16 Rev 10 06 2021 USM Go ...
Page 267: ...USM Go Rev 10 06 2021 10 1 Specifications 10 ...
Page 278: ...10 12 Rev 10 06 2021 USM Go ...