Pagina 6.26
6.9.
“THE EQUIPMENT IS BLOCKED (COM)” MESSAGE ON
DISPLAY
This error message can be caused by loss of communication between the lower and upper
assemblies.
The error is displayed by a massage and stored in the “
Errors Log
” section.
To reinstate communications between the lower and upper assemblies:
(1)
Check that cables
CU69
and
CU63
are correct, using the Test Box Excite and replace if
necessary;
(2)
Try replacing the brake board and the display board in turn, and check whether the
communication works.
If you replace Display board/Brake board, check that its SW version is updated,
otherwise install the last version.