![WAVE EXCITE CLASS & TREND Service Maintenance Manual Download Page 186](http://html1.mh-extra.com/html/wave-excite/class-and-trend/class-and-trend_service-maintenance-manual_877550186.webp)
CARDIO WAVE EXCITE Class & Trend: Service & maintenance manual- rev.2.0
Pagina 9.34
9.5.3.3.
COM.Fault
This is a counter that display the number of errors in serial communication between the upper and
the lower kit. This counter increase each time a communication fault is detected.
To reset this counter, press
RESET
key on the side of the value.
To correctly display the parameter values, you need to load them from the low kit, using
the “Read from low kit” function.
For every error generated by the machine, the error history log records the information in the table
below:
•
Counter
•
Error code
•
Current when the error occurred
•
Distance in km when the error occurred
The correspondence between error codes and descriptions is given in the “Service & maintenance
manual”.
Press
HOME
to confirm and save,
FORWARD
or
BACK
to scroll the pages.