CARDIO WAVE EXCITE Class & Trend: Service & maintenance manual- rev.2.0
Pagina 7.32
7.15.
DISASSEMBLING THE BRAKE WINDING
Figure 7.15-1
Carry out the operations described in
paragraphs: 7.7. “Disassembling the guards”,
only the anterior guard.
1.
Unplug the two faston connectors
(a)
from
the winding.
2.
Unplug the connector circled in the figure at
left.
Figure 7.15-2
3.
Back off the 2 screws
(b)
using an 17mm
wrench, supporting the entire brake winding
group
(c)
, to avoid damaging the copper disk.
4.
Remove the brake winding assembly
To reassemble the brake winding, carry out the
above steps in reverse order.
After completing the reassembly, adjust
the position of the winding group as
described in paragraph 8.3. “Brake
winding position”.