21.8
Pump parameters
Setting parameters
To set a parameter to a new value:
•
Type a value into the field or click on the check box (depending on the parameter type)
•
Click 'set' to store the new value or click 'refresh' to cancel the change
•
Up to 10 parameters are displayed per page. Use the < and > buttons to navigate between pages
The following parameters can be set by the user.
Param
No.
EDS Values
Read/Write
Default
Value
(EDS)
Parameter
Range
Comments
1
SetFlowCal
Write
15120
1 -
2147483647
Use to set the tube flow
calibration value. The flow
calibration value is set in
μL. For more information
on flow calibration
see
2
SetSpeed
Write
100
1 - 2200
Speed is set in Deci RPM.
Max speed depends on
head type.See
.
3
SetSpeedLimit
Write
2200
1 - 2200
Speed is set in Deci RPM.
Max speed depends on
head type. See
.
4
SetFailsafeSpeed
Write
100
1 - 2200
If the failsafe is enabled
the pump will run
continuously at the
selected speed in the
event of a
communications loss.
80