22.1
Calibrate the pump for 4-20 mA control (Un only)
This feature is only available in the Un model.
The pump must be stopped before attempting to calibrate the 4-20 mA values. High and
low signals must be within range. If the signal sent is out of range you will not be able to
set the signal input value and progress to the next step in the process.
Select
MODE
.
Using the
+/-
keys scroll to
Analog 4-20 mA
and press
CALIBRATE 4-20 mA
.
Choose whether to enter the current values manually via the keypad, or whether to apply
the current signals electrically to the analog input.
m-qdos-allmodels-en-07
88