Recipe name
The recipe name gives the user a convenient way to identify a recipe.
Maximum 12 characters. (A-Z, 0-9).
•
Using the
∧
/
∨
keys, enter a character.
•
Press
NEXT
to confirm your character and to move onto the next. Press
PREVIOUS
to move
the cursor back one character.
•
Move the cursor to the end or beginning of the entry field using
NEXT
or
PREVIOUS
. When the
cursor is at the beginning or end of the entry field, press
FINISH
to store the entry.
Volume
This sets the target dispense amount.
•
Minimum = 0.1000 millilitres
•
Maximum = 99999.9 millilitres
Speed
Rotor speed.
Table 18 - Maximum pump speed
530Pn/PnN
220 rpm
Excessive speed may cause splashing or foaming.
Anti-drip
If dripping occurs after the fill has completed, increase anti-drip to create "suck back" by momentarily
reversing the direction of pumphead. Anti-drip is measured by the number of reverse steps 0-10.
Anti-drip values are whole numbers between 0 and 10 where 10 is one complete reverse revolution of
the rotor and 0 is no reverse revolution of the rotor.
When using anti-drip, prime the pump before starting each new batch. This will compensate for the
volume of fluid that has been retracted due to anti-drip.
Note: To reduce dripping always use an appropriate filling needle and ensure it is perfectly vertical.
Start delay (Recipe)
Sets time delay between start signal and dose start.
Refer to "Dispense time delays diagram" on page 110.
End delay (Recipe)
Sets time delay between pumphead stop and dose complete signal.
Refer to "Dispense time delays diagram" on page 110.
Start ramp
This will set the rate of acceleration when the pump starts.
Can be set to a value between 1 and 5.
1 is the fastest acceleration, 5 is the slowest.
Note: Start ramp is not included in flow calibration.
108