March 14, 2014, 715004387 Rev. B
v
with. Always follow Good Laboratory Practice, and consult your organization’s
standard operating procedures.
Safety hazard symbol notice
Documentation needs to be consulted in all cases where the
symbol is
used to find out the nature of the potential hazard and any actions which have
to be taken.
Considerations specific to the Xevo TQD
Power cord replacement hazard
Solvent leakage hazard
The source exhaust system is designed to be robust and leak-tight. Waters
recommends you perform a hazard analysis, assuming a maximum leak into
the laboratory atmosphere of 10% LC eluate.
Warning:
To avoid electric shock, use the SVT-type power cord in the
United States and HAR-type (or better) in Europe. The main power cord
must only be replaced with one of adequate rating. For information
regarding what cord to use in other countries, contact your local Waters
distributor.
Warning:
• To confirm the integrity of the source exhaust system, renew
the source O-rings at intervals not exceeding one year.
• To avoid chemical degradation of the source O-rings, which can
withstand exposure only to certain solvents (see
,
determine whether any solvents you use that are not listed are
chemically compatible with the composition of the O-rings.
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...