4 Maintenance Procedures
176
March 14, 2014, 715004387 Rev. B
4. Dispose of all the seals in accordance with local environmental
regulations.
Fitting the new source enclosure and probe adjuster assembly
seals
Required materials
•
Chemical-resistant, powder-free gloves
•
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
•
New seals
Warning:
To avoid spreading contamination, dispose of the
seals in accordance with local environmental regulations.
The seals can be contaminated with biohazardous or toxic
materials.
TP03164
Source enclosure seal
Nebulizer gas seal
Desolvation gas seal