4 Maintenance Procedures
102
March 14, 2014, 715004387 Rev. B
8. To start the LC flow, follow the action below for your software:
Gas ballasting the roughing pump
The roughing pump draws large quantities of solvent vapors. The vapors tend
to condense in the pump oil, diminishing pumping efficiency. Gas ballasting
purges condensed contaminants from the oil.
Note:
This procedure is not required for an oil-free roughing pump.
Roughing pump:
Software
Action
MassLynx
In the instrument console, click Start Flow
.
UNIFI
On the System Console tool bar, click Start Flow
.
TP02689
Oil filler plug
Oil-level sight glass
Drain plug
Gas ballast valve
Exhaust port flange