Combined APPI/APCI source
3-13
APPI lamp drive assembly inside the source enclosure:
Installing the combined APPI/APCI source
Required material
Chemical-resistant, powder-free gloves
To install the combined APPI/APCI source:
1. Prepare the instrument for working on the source (see
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, ensure that the instrument is
prepared for working on the source before commencing this procedure.
TP03201
APPI lamp drive
assembly
Source enclosure
UV lamp and
repeller electrode
IonSABRE II probe