4-94
Maintenance Procedures
4. Dispose of all the seals in accordance with local environmental
regulations.
Fitting the new source enclosure and probe adjuster assembly
seals
Required materials
•
Chemical-resistant, powder-free gloves
•
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
•
New seals
To fit the new source enclosure and probe adjuster assembly probe seals:
1. Ensure that all the grooves for seals are free from dirt and debris.
Tip:
If contamination is present, use 1:1 methanol/water, applied to a
lint-free cloth, to carefully clean the grooves.
2. Fit the new source enclosure seal to the source enclosure.
Tip:
Start by feeding the seal into the groove at the bottom right-hand
corner, and then working around the seal in a counterclockwise
direction.
3. Fit the following new seals to the source enclosure:
•
Nebulizer gas seal
Warning:
The seals can be contaminated with biohazardous
and/or toxic materials. Dispose of them according to local
environmental regulations.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Caution:
Ensure that the tails of the source enclosure seals are
correctly located in the groove when fitting them to the source
enclosure.
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...