Maintenance procedures
This section provides the maintenance guidelines and procedures necessary to maintain the
mass spectrometer's performance.
Keep to a maintenance schedule, and perform maintenance as required and described in this
section.
4.1 Maintenance schedule
The following table lists periodic maintenance schedules that ensure optimum instrument
performance.
Table 4–1: Maintenance schedule:
Procedure
Frequency
For information...
Clean the instrument case.
As required.
Empty the nitrogen exhaust
trap bottle.
Check daily, empty as required.
Clean the source
components.
When source components are
visibly fouled, the background or
high-peak contaminants are
unacceptably high, or sensitivity
decreases to unacceptable levels.
Replace the ESI probe tip.
When sensitivity decreases to
unacceptable levels, or if blocked
or damaged.
Replace the probe
assembly.
When sensitivity decreases to
unacceptable levels or signal is
unstable due to inconsistent
sample flow.
Clean the APCI probe tip.
(Options using the APCI
probe adaptor only.)
When sensitivity decreases to
unacceptable levels or when
significant chemical interference
is present.
.